Conversion of mammographic images to appear with the noise and sharpness characteristics of a different detector and x‐ray system
Authors:
- Alistair MackenzieAMAlistair Mackenzie
- David R. DanceDDDavid R. Dance
- Adam WorkmanAWAdam Workman
- Mary YipMYMary Yip
- Kevin WellsKWKevin Wells
- Kenneth C. YoungKYKenneth C. Young
- 23 April 2012
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 39 (5) , 2721-2734
Abstract
Purpose:: Undertaking observer studies to compare imaging technology using clinical radiological images is challenging due to patient variability. To achieve a significant result, a large number of patients would be required to compare cancer detection rates for different image detectors and systems. The aim of this work was to create a methodology where only one set of images is collected on one particular imaging system. These images are then converted to appear as if they had been acquired on a different detector and x‐ray system. Therefore, the effect of a wide range of digital detectors on cancer detection or diagnosis can be examined without the need for multiple patient exposures.Methods:: Three detectors and x‐ray systems [Hologic Selenia (ASE), GE Essential (CSI), Carestream CR (CR)] were characterized in terms of signal transfer properties, noise power spectra (NPS), modulation transfer function, and grid properties. The contributions of the three noise sources (electronic, quantum, and structure noise) to the NPS were calculated by fitting a quadratic polynomial at each spatial frequency of the NPS against air kerma. A methodology was developed to degrade the images to have the characteristics of a different (target) imaging system. The simulated images were created by first linearizing the original images such that the pixel values were equivalent to the air kerma incident at the detector. The linearized image was then blurred to match the sharpness characteristics of the target detector. Noise was then added to the blurred image to correct for differences between the detectors and any required change in dose. The electronic, quantum, and structure noise were added appropriate to the air kerma selected for the simulated image and thus ensuring that the noise in the simulated image had the same magnitude and correlation as the target image. A correction was also made for differences in primary grid transmission, scatter, and veiling glare. The method was validated by acquiring images of a CDMAM contrast detail test object (Artinis, The Netherlands) at five different doses for the three systems. The ASE CDMAM images were then converted to appear with the imaging characteristics of target CR and CSI detectors.Results:: The measured threshold gold thicknesses of the simulated and target CDMAM images were closely matched at normal dose level and the average differences across the range of detail diameters were −4% and 0% for the CR and CSI systems, respectively. The conversion was successful for images acquired over a wide dose range. The average difference between simulated and target images for a given dose was a maximum of 11%.Conclusions:: The validation shows that the image quality of a digital mammography image obtained with a particular system can be degraded, in terms of noise magnitude and color, sharpness, and contrast to account for differences in the detector and antiscatter grid. Potentially, this is a powerful tool for observer studies, as a range of image qualities can be examined by modifying an image set obtained at a single (better) image quality thus removing the patient variability when comparing systems.Keywords
This publication has 28 references indexed in Scilit:
- Image quality assessment in digital mammography: part I. Technical characterization of the systemsPhysics in Medicine & Biology, 2011
- Simulation of dose reduction in tomosynthesisMedical Physics, 2009
- The effect of scatter and glare on image quality in contrast‐enhanced breast imaging using an full‐field flat panel detectorMedical Physics, 2009
- A Technique for Simulating the Effect of Dose Reduction on Image Quality in Digital Chest RadiographyJournal of Digital Imaging, 2008
- Retrospective analysis of a detector fault for a full field digital mammography systemPhysics in Medicine & Biology, 2006
- Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer ScreeningNew England Journal of Medicine, 2005
- Method of simulating dose reduction for digital radiographic systemsRadiation Protection Dosimetry, 2005
- Screen-Film Mammography versus Full-Field Digital Mammography with Soft-Copy Reading: Randomized Trial in a Population-based Screening Program—The Oslo II StudyRadiology, 2004
- An adaptive algorithm for the detection of microcalcifications in simulated low-dose mammographyPhysics in Medicine & Biology, 2003
- Signal, noise and SNR transfer properties of computed radiographyPhysics in Medicine & Biology, 1993