Method of simulating dose reduction for digital radiographic systems
Authors:
- Magnus BåthMBMagnus Båth
- Markus HåkanssonMHMarkus Håkansson
- Anders TingbergATAnders Tingberg
- Lars Gunnar MånssonLMLars Gunnar Månsson
- 17 May 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 114 (1-3) , 253-259
Abstract
The optimisation of image quality vs. radiation dose is an important task in medical imaging. To obtain maximum validity of the optimisation, it must be based on clinical images. Images at different dose levels can then either be obtained by collecting patient images at the different dose levels sought to investigate—including additional exposures and permission from an ethical committee—or by manipulating images to simulate different dose levels. The aim of the present work was to develop a method of simulating dose reduction for digital radiographic systems. The method uses information about the detective quantum efficiency and noise power spectrum at the original and simulated dose levels to create an image containing filtered noise. When added to the original image this results in an image with noise which, in terms of frequency content, agrees with the noise present in an image collected at the simulated dose level. To increase the validity, the method takes local dose variations in the original image into account. The method was tested on a computed radiography system and was shown to produce images with noise behaviour similar to that of images actually collected at the simulated dose levels. The method can, therefore, be used to modify an image collected at one dose level so that it simulates an image of the same object collected at any lower dose level.Keywords
This publication has 20 references indexed in Scilit:
- Nodule detection in digital chest radiography: effect of system noiseRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: introduction to the RADIUS chest trialRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: part of image background acting as pure noiseRadiation Protection Dosimetry, 2005
- Comparison of two methods for evaluation of image quality of lumbar spine radiographs2004
- How do radiographic techniques affect mass lesion detection performance in digital mammography?2004
- Relative impact of detector noise and anatomical structure on lung nodule detection2004
- Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiographyEuropean Radiology, 2003
- Human observer detection experiments with mammograms and power‐law noiseMedical Physics, 2001
- Detection of Subtle Lung Nodules: Relative Influence of Quantum and Anatomic Noise on Chest RadiographsRadiology, 1999
- Estimation of the noisy component of anatomical backgroundsMedical Physics, 1999