Far‐Field R‐Wave Reduction with a Novel Lead Design: Experimental and Human Results
Authors:
- 11 August 2005
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 28 (8) , 782-788
Abstract
Introduction: The purpose of this study was to examine a bipolar screw‐in lead (NL), specially designed to reduce unwanted far‐field R‐wave (FFRW) signal detection in an acute human setting. The results were compared with animal experiments. Methods: The newly designed lead with a center‐to‐center distance between the anode and cathode electrodes of 3.23 mm, corresponding to an inter‐electrode spacing of 1.1 mm was implanted in nine canines with a follow‐up of 6 months. Sensing of P waves, FFRW signals, pacing threshold, and impedance was measured at regular intervals. As a result of the positive outcome with the animal study, an acute human experiment was performed. In patients scheduled for conventional dual chamber pacemaker implantation, the NL was compared to a Tendril® Model 1388T bipolar screw‐in lead (St. Jude Medical, CRMD, Sylmar, CA). Results: Utilizing a tip‐to‐ring distance of 1.1 mm, the optimum P wave to FFRW ratio was found in animal experiments. In the acute human tests in 15 patients, the mean P‐wave voltage of the 1388T lead of 3.30 ± 1.54 mV was slightly larger than that of the NL, at 2.55 ± 1.11 mV, but did not differ significantly (P = 0.13). The FFRW voltage of the 1388T lead was 0.62 ± 0.37 mV and was significantly greater from that of the NL, at 0.10 ± 0.08 mV (P < 0.0001). Pacing thresholds and pacing impedances were comparable. Conclusion: Animal testing results were reproducible in the acute human test setting. The lead reduced the paced FFRW signal amplitudes significantly, allowing for high atrial sensitivity settings but without sensing the FFRW. A robust P‐wave signal could be retained.Keywords
This publication has 13 references indexed in Scilit:
- Far‐Field R Wave Oversensing in Dual Chamber Pacemakers Designed for Atrial Arrhythmia Management:Pacing and Clinical Electrophysiology, 2004
- Atrial pacing for prevention of atrial fibrillation: assessment of simultaneously implemented algorithmsEP Europace, 2004
- Accuracy of Atrial Tachyarrhythmia Detection in Implantable Devices with Arrhythmia TherapiesPacing and Clinical Electrophysiology, 2004
- Accuracy of Mode Switch Algorithms for Detection of Atrial TachyarrhythmiasJournal of Cardiovascular Electrophysiology, 2004
- Combined Efficacy of Atrial Septal Lead Placement and Atrial Pacing Algorithms for Prevention of Paroxysmal Atrial TachyarrhythmiaJournal of Cardiovascular Electrophysiology, 2003
- Impact and Prevention of Far‐Field Sensing in Fallback Mode SwitchesPacing and Clinical Electrophysiology, 2003
- Atrial Lead Implantation During Atrial Flutter or Fibrillation?Pacing and Clinical Electrophysiology, 1998
- Analysis of Atrial Sensed Far‐Field Ventricnlar Signals: A ReassessmentPacing and Clinical Electrophysiology, 1997
- Natural history of sinus node disease treated with atrial pacing in 213 patients: Implications for selection of stimulation modeJACC, 1992
- Far‐Field QRS Complex Sensing Via the Atrial Pacemaker Lead. II. Prevalence, Clinical Significance and Possibility of Intraoperative Prediction in DDD Pacing.Pacing and Clinical Electrophysiology, 1988