Segmental acetabular rim defects, bone loss, oversizing, and press fit cup in total hip arthroplasty evaluated with a probabilistic finite element analysis
- 23 December 2016
- journal article
- research article
- Published by Springer Nature in International Orthopaedics
- Vol. 41 (8) , 1527-1533
Abstract
Purpose Management of segmental rim defects and bone mineral density (BMD) loss in the elderly prior to total hip replacement is unclear within classification systems for acetabular bone loss. In this study, our objectives were (1) to understand how a reduction in BMD in the elderly affects the oversizing of a press-fit cup for primary fixation and (2) to evaluate whether the location of the segmental defect affected cup fixation. Methods A finite element (FE) model was used to simulate and evaluate cup insertion and fixation in the context of segmental rim defects. We focused on the distribution of patients over age 70 and used BMD (estimated from CT) as a proxy for aging’s implications on THR and used probabilistic FE analysis to understand how BMD loss affects oversizing of a press-fit cup. Results A cup oversized by 1.10 ± 0.28 mm provides sufficient fixation and lower stresses at the cup-bone interface for elderly patients. Defects in the anterior column and posterior column both required the same mean insertion force for cup seating of 84% (taken as an average of 2 anterior column and 2 posterior column defects) compared to the control configuration, which was 5% greater than the insertion force for a superior rim defect and 12% greater than the insertion force for an inferior rim defect. Conclusions A defect along the superior or inferior rim had a minimal effect on cup fixation, while a defect in the columns created cup instability and increased stress at the defect location.Keywords
Funding Information
- Ariello Cocommo Biomechanics Fund (001-N/A)
This publication has 24 references indexed in Scilit:
- Revision rates after total joint replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Challenges in modeling total knee arthroplasty and total hip replacementProcedia IUTAM, 2011
- Reliability of radiographic signs for acetabular retroversionInternational Orthopaedics, 2010
- The Effect of Under-Reaming on the Cup/Bone Interface of a Press Fit Hip ReplacementJournal of Biomechanical Engineering, 2010
- Projections of Primary and Revision Hip and Knee Arthroplasty in the United States from 2005 to 2030Journal of Bone and Joint Surgery, 2007
- Subject-specific finite element models of long bones: An in vitro evaluation of the overall accuracyJournal of Biomechanics, 2006
- The role of the anterior column of the acetabulum on pelvic stability: a biomechanical studyInjury, 1998
- The role of the anterior column of the acetabulum on pelvic stability: a biomechanical studyInjury, 1998
- JUMBO CUPS AND MORSELIZED GRAFTOrthopedic Clinics of North America, 1998
- Predicting the compressive mechanical behavior of boneJournal of Biomechanics, 1994