Heat production during pulmonary artery sealing with energy vessel-sealing devices in a swine model
- 5 November 2020
- journal article
- conference paper
- Published by Oxford University Press (OUP) in Interactive CardioVascular and Thoracic Surgery
- Vol. 31 (6) , 847-852
Abstract
OBJECTIVES: Energy vessel-sealing devices are being increasingly utilized to seal pulmonary artery (PA) branches during lobectomy. Heat from these devices can potentially injure surrounding tissues. We evaluated heat production from devices in a live animal model. METHODS: PA branches were sealed in pigs with 4 energy vessel-sealing devices: 2 ultrasonic (US), 1 advanced bipolar or 1 mixed US and bipolar (mixed) device. Thermocouples were implanted in tissue surrounding the PA branch being sealed to measure tissue temperature. A thermal camera measured the sealing site and the temperatures of the instruments. Pathological analysis was performed on PA stumps to identify thermal damage. RESULTS: A total of 37 PA branches were sealed in 4 pigs. Maximum tissue heat measured by the thermocouples for the 2 US, advanced bipolar and mixed devices was 42, 39, 42 and 46°C, respectively. The mean tissue temperatures at the site of the sealing measured with the thermal camera were 78, 75, 70 and 82°C (P = 0.834) and the mean instrument blade temperatures were 224, 195, 83 and 170°C (P = 0.000005) for the 2 US, advanced bipolar and mixed devices, respectively. The mean diameter of the region with tissue reaching 60°C or more measured with the thermal camera was between 4 and 6 mm for the 4 devices (P = 0.941). On pathological analysis, PA stumps had either thermal damage on the adventitia and external media (26/37) or transmural damage (11/37) at 1 mm from sealed site. CONCLUSIONS: A 3-mm safety margin between the instrument blades and vital structures is recommended. Instrument blades can reach high temperatures that may cause tissue damage.Keywords
Funding Information
- Marcel and Rollande Gosselin Chair in Thoracic Surgical Oncology
This publication has 21 references indexed in Scilit:
- Pulmonary Artery Sealing Using the HARMONIC ACE+ Shears for Video-Assisted Thoracoscopic Surgery LobectomyThe Annals of Thoracic Surgery, 2015
- Surgical Energy-Based Device Injuries and Fatalities Reported to the Food and Drug AdministrationJournal of the American College of Surgeons, 2015
- Harmonic Scalpel and Electrothermal Bipolar Vessel Sealing System in Head and Neck SurgeryAnnals of Otology, Rhinology & Laryngology, 2015
- Pilot study of pulmonary arterial branch sealing using energy devices in an ex vivo modelThe Journal of Thoracic and Cardiovascular Surgery, 2014
- Usefulness of vessel-sealing devices for ≤7 mm diameter vessels: a randomized controlled trial for human thoracoscopic lobectomy in primary lung cancerInteractive CardioVascular and Thoracic Surgery, 2014
- Sealing of pulmonary arteries with LigaSure: In vivo and ex vivo examinationsThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Preliminary results of anatomic lung resection using energy-based tissue and vessel coagulative fusion technologyThe Journal of Thoracic and Cardiovascular Surgery, 2010
- Tissue Response to Surgical Energy DevicesUrology, 2008
- Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT)Surgical Endoscopy, 2007
- Image-guided Radiofrequency Tumor Ablation: Challenges and Opportunities—Part IJournal of Vascular and Interventional Radiology, 2001