Hepatic bipolar radiofrequency ablation using perfused-cooled electrodes: a comparative study in theex vivobovine liver
Authors:
- J K HanJHJ K Han
- S H KimSKS H Kim
- S H ChoiSCS H Choi
- B I ChoiBCB I Choi
- 1 November 2004
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 77 (923) , 944-949
Abstract
The purpose of this paper was to demonstrate the efficacy of the dual probe bipolar radiofrequency (RF) system with the perfused-cooled electrodes inducing coagulation necrosis in the ex vivo bovine liver. The perfused-cooled electrode that allows simultaneous internal cooling and interstitial hypertonic saline perfusion has been developed for RF ablation (RFA). RF was applied to excised bovine liver in a bipolar mode at 150 W using a 200 W generator with two perfused-cooled electrodes for 10 min. After placing the electrodes at 3 cm spacing in the explanted liver, 45 ablation zones were created with three different regimens: Group A, using both intraelectrode cooling and interstitial perfusion; group B, using only the intraelectrode cooling; and group C, using only interstitial perfusion. In groups A and C, RFA was performed with the infusion of 6% hypertonic saline at the rate of 2 ml min21. During RFA, we measured the tissue temperature at the midpoint between the two electrodes. The dimensions of the ablation zones and the changes in impedance, currents and liver temperature during RFA were compared in these three groups. The mean tissue impedance during RFA in group A (56.7¡21.7 V) and group C (56.9¡20.6 V) was significantly lower than group B (112¡19.7 V) (p,0.001). The mean current was higher in group A (1765¡128 mA) than groups B (760¡321 mA) and C (1298¡349 mA) (p,0.05). In addition, the shortest vertical diameter of coagulation necrosis was greater in groups A (4.9¡0.5 cm) and C (4.6¡0.7 cm) than in group B (3.5¡0.4 cm) (p,0.05). The temperature at the mid-point between the two probes was higher in group A than other groups: 99˚C in group A, 88.9˚C in group B, and 94.3˚C in group C (p.0.05). The ratios of the diameter of the long-axis to the diameter of the vertical- axis of groups A, B and C were 1.1¡0.1, 1.2¡0.1, and 1.1¡0.2, respectively (p,0.05). Bipolar RFA using intraelectrode cooling and the interstitial saline perfusion simultaneously produced ablation zones significantly larger than the area produced by only one measure.Keywords
This publication has 36 references indexed in Scilit:
- Percutaneous ablation of liver tumoursBest Practice & Research Clinical Gastroenterology, 2004
- Risk factors for the local recurrence of hepatocellular carcinoma after a single session of percutaneous radiofrequency ablationJournal of Gastroenterology, 2003
- Risk factors for local recurrence of small hepatocellular carcinoma tumors after a single session, single application of percutaneous radiofrequency ablationCancer, 2003
- Percutaneous Local Ablative Therapy for Hepatocellular CarcinomaAnnals of Surgery, 2003
- A Comparative Experimental Study of the In-vitro Efficiency of Hypertonic Saline-Enhanced Hepatic Bipolar and Monopolar Radiofrequency AblationKorean Journal of Radiology, 2003
- Overlapping Ablation Using a Coaxial Radiofrequency Electrode and Multiple Cannulae System: Experimental Study in ex-Vivo Bovine LiverKorean Journal of Radiology, 2003
- Percutaneous Radiofrequency Thermal Ablation with Hypertonic Saline Injection: In Vivo Study in a Rabbit Liver ModelKorean Journal of Radiology, 2003
- Radiofrequency tumor ablation: principles and techniquesEuropean Journal of Ultrasound, 2001
- MR Imaging-Histopathologic Correlation of Radiofrequency Thermal Ablation Lesion in a Rabbit Liver Model: Observation during Acute and Chronic StagesKorean Journal of Radiology, 2001
- Radiofrequency Ablation of Porcine Liver In VivoAnnals of Surgery, 1998