Estimation and comparison of cancer incidence risk using two algorithms in multi-slice CT systems for patients undergoing coronary CT angiography
- 21 October 2025
- journal article
- Published by Taylor & Francis in International Journal of Radiation Biology
- Vol. 102 (1) , 75-84
Abstract
The increasing use of Coronary CT Angiography (CCTA) raises concerns regarding radiation exposure and its associated cancer risks. This study aims to evaluate the effectiveness of the Siemens CARE kV algorithm in reducing radiation doses and cancer incidence risk compared to the Automatic Exposure Control (AEC) algorithm in adult patients undergoing CCTA. This retrospective study involved a cohort of 101 patients, comprising 55 women and 46 men, who underwent Siemens single-source 1 × 64-slice Multi-Detector CT (Somatom Definition AS) for CCTA. Demographic data and radiation dose parameters, including Computed Tomography Dose Index (CTDIvol) and Dose-Length Product (DLP) were recorded. Effective and organ doses were calculated using ImPACT software, and Lifetime Attributable Risk (LAR) was estimated based on BEIR VII data. The Effective Doses (EDs) for AEC and CARE kV algorithms were 9.59 ± 3.84 mSv and 8.04 ± 3.99 mSv, respectively. The cancer incidence risk was significantly lower for the CARE kV algorithm, with LAR estimates of 7.27 ± 4.01 per 100,000 populations compared to 13.88 ± 10.76 for AEC. The CARE kV algorithm demonstrated a 27.07% reduction in LAR. The findings suggest that the CARE kV algorithm provides a significant reduction in radiation exposure and associated cancer risks compared to the AEC algorithm. These results support the optimization of CT protocols to enhance patient safety while maintaining diagnostic efficacy.Keywords
Funding Information
- Urmia University of Medical Sciences
- Urmia University of Medical Sciences (11420)
This publication has 35 references indexed in Scilit:
- A phantom based evaluation of the dose prediction and effects in treatment plans, when calculating on a direct density CT reconstructionJournal of Applied Clinical Medical Physics, 2020
- Risk of low-dose radiation and the BEIR VII report: A critical review of what it does and doesn’t sayPhysica Medica, 2017
- Abdominal imaging dose in radiology and radiotherapy – Phantom point dose measurements, effective dose and secondary cancer riskPhysica Medica, 2017
- Achieving consistent image quality and overall radiation dose reduction for coronary CT angiography with body mass index-dependent tube voltage and tube current selectionClinical Radiology, 2014
- ALARA (as low as reasonably achievable) CT 2011—executive summaryPediatric Radiology, 2011
- Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007Archives of internal medicine (1960), 2009
- Radiation Dose From Cardiac Computed Tomography Before and After Implementation of Radiation Dose–Reduction TechniquesJAMA, 2009
- Dose reduction by automatic exposure control in multidetector computed tomography: comparison between measurement and calculationEuropean Radiology, 2008
- Strategies for reduction of radiation dose in cardiac multislice CTEuropean Radiology, 2007
- Aorto-iliac multidetector-row CT angiography with low kV settings: improved vessel enhancement and simultaneous reduction of radiation doseEuropean Radiology, 2004