Experimental determination of human peripheral nerve stimulation thresholds in a 3‐axis planar gradient system
- 12 June 2009
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 62 (3) , 763-770
Abstract
In MRI, strong, rapidly switched gradient fields are desirable because they can be used to reduce imaging time, obtain images with better resolution, or improve image signal‐to‐noise ratios. Improvements in gradient strength can be made by either increasing the gradient amplifier strength or by enhancing gradient efficiency. Unfortunately, many MRI pulse sequences, in combination with high‐performance amplifiers and existing gradient hardware, can cause peripheral nerve stimulation (PNS). This makes improvements in gradient amplifiers ineffective at increasing safely usable gradient strength. Customized gradient coils are one way to achieve significant improvements in gradient performance. One specific gradient configuration, a planar gradient system, promises improved gradient strength and switching time for cardiac imaging. The PNS thresholds for planar gradients were characterized through human stimulation experiments on all three gradient axes. The specialized gradient was shown to have significantly higher stimulation thresholds than traditional cylindrical designs (y‐axis SRmin = 210 ± 18 mT/m/ms and ΔGmin = 133 ± 13 mT/m; x‐axis SRmin = 222 ± 24 mT/m/ms and ΔGmin = 147 ± 17 mT/m; z‐axis SRmin = 252 ± 26 mT/m/ms and ΔGmin = 218 ± 26 mT/m). This system could be operated at gradient strengths 2 to 3 times higher than cylindrical configurations without causing stimulation. Magn Reson Med, 2009.Keywords
This publication has 20 references indexed in Scilit:
- Local planar gradients with order‐of‐magnitude strength and speed advantageMagnetic Resonance in Medicine, 2007
- Accuracy Limitations of Chronaxie ValuesIEEE Transactions on Biomedical Engineering, 2004
- Peripheral nerve stimulation properties of head and body gradient coils of various sizesMagnetic Resonance in Medicine, 2003
- Simple anatomical measurements do not correlate significantly to individual peripheral nerve stimulation thresholds as measured in MRI gradient coilsJournal of Magnetic Resonance Imaging, 2003
- Comparison of the threshold for peripheral nerve stimulation during gradient switching in whole body MR systemsJournal of Magnetic Resonance Imaging, 2002
- A comparison between human magnetostimulation thresholds in whole‐body and head/neck gradient coilsMagnetic Resonance in Medicine, 2001
- Simple linear formulation for magnetostimulation specific to MRI gradient coilsMagnetic Resonance in Medicine, 2001
- Physiological Effects of Fast Oscillating Magnetic Field GradientsJournal of Computer Assisted Tomography, 1991
- Considerations of quasi-stationarity in electrophysiological systemsBulletin of Mathematical Biology, 1967
- The effect upon the threshold for nervous excitation of the length of nerve exposed, and the angle between current and nerveThe Journal of Physiology, 1927