Abstract
Total shoulder arthroplasty (TSA) is an effective treatment for glenohumeral (GH) osteoarthritis. However, it still suffers from a substantial rate of mechanical failure, which may be related to cyclic off-center loading of the humeral head on the glenoid. In this work, we present the design and evaluation of a GH joint robotic simulator developed to study GH translations. This five-degree-of-freedom robot was designed to replicate the rotations (±40 deg, accuracy 0.5 deg) and three-dimensional (3D) forces (up to 2 kN, with a 1% error settling time of 0.6 s) that the humeral implant exerts on the glenoid implant. We tested the performances of the simulator using force patterns measured in real patients. Moreover, we evaluated the effect of different orientations of the glenoid implant on joint stability. When simulating realistic dynamic forces and implant orientations, the simulator was able to reproduce stable behavior by measuring the translations of the humeral head of less than 24 mm with respect to the glenoid implant. Simulation with quasi-static forces showed dislocation in extreme ranges of implant orientation. The robotic GH simulator presented here was able to reproduce physiological GH forces and may therefore be used to further evaluate the effects of glenoid implant design and orientation on joint stability.Keywords
Funding Information
- Swiss National Science Foundation (CR32I2_162766)
This publication has 31 references indexed in Scilit:
- Biomechanical consequences of humeral component malpositioning after anatomical total shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2010
- Prevalence and projections of total shoulder and elbow arthroplasty in the United States to 2015Journal of Shoulder and Elbow Surgery, 2010
- Humeral Head Posterior Subluxation on CT Scan: Validation and Comparison of 2 Methods of MeasurementTechniques in Shoulder & Elbow Surgery, 2010
- The effect of component positioning on intrinsic stability of the reverse shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2010
- Glenoid loosening after total shoulder arthroplasty: An in vitro CT‐scan studyJournal of Orthopaedic Research, 2009
- Total shoulder arthroplasty: Downward inclination of the glenoid component to balance supraspinatus deficiencyJournal of Shoulder and Elbow Surgery, 2009
- In vitro and finite element analysis of glenoid bone/baseplate interaction in the reverse shoulder designJournal of Shoulder and Elbow Surgery, 2008
- The effect of glenoid inclination on superior humeral head migrationJournal of Shoulder and Elbow Surgery, 2003
- A finite element musculoskeletal model of the shoulder mechanismJournal of Biomechanics, 1994