Comparison between TG-51 and TG-21: Calibration of photon and electron beams in water using cylindrical chambers
Open Access
- 1 June 2000
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 1 (3) , 108-115
Abstract
A new calibration protocol, developed by the AAPM Task Group 51 (TG‐51) to replace the TG‐21 protocol, is based on an absorbed‐dose to water standard and calibration factor (ND,w), while the TG‐21 protocol is based on an exposure (or air‐kerma) standard and calibration factor (Nx). Because of differences between these standards and the two protocols, the results of clinical reference dosimetry based on TG‐51 may be somewhat different from those based on TG‐21. The Radiological Physics Center has conducted a systematic comparison between the two protocols, in which photon and electron beam outputs following both protocols were compared under identical conditions. Cylindrical chambers used in this study were selected from the list given in the TG‐51 report, covering the majority of current manufacturers. Measured ratios between absorbed‐dose and air‐kerma calibration factors, derived from the standards traceable to the NIST, were compared with calculated values using the TG‐21 protocol. The comparison suggests that there is roughly a 1% discrepancy between measured and calculated ratios. This discrepancy may provide a reasonable measure of possible changes between the absorbed‐dose to water determined by TG‐51 and that determined by TG‐21 for photon beam calibrations. The typical change in a 6 MV photon beam calibration following the implementation of the TG‐51 protocol was about 1%, regardless of the chamber used, and the change was somewhat smaller for an 18 MV photon beam. On the other hand, the results for 9 and 16 MeV electron beams show larger changes up to 2%, perhaps because of the updated electron stopping power data used for the TG‐51 protocol, in addition to the inherent 1% discrepancy presented in the calibration factors. The results also indicate that the changes may be dependent on the electron energy.Keywords
Funding Information
- National Cancer Institute
- Public Health Service (CA 10953)
- Department of Health and Human Services
This publication has 5 references indexed in Scilit:
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- Clinical electron‐beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25Medical Physics, 1991
- Clarification of the AAPM Task Group 21 protocolMedical Physics, 1986
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983