Ablation catheter orientation: In vitro effects on lesion size and in vivo analysis during PVI for atrial fibrillation
- 28 October 2020
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 43 (12) , 1554-1563
Abstract
Background: Catheter‐delivered radiofrequency (RF) lesion formation is a complex phenomenon, and few studies have explored the effect of catheter orientation on lesion size and catheter orientation behavior during pulmonary vein isolation (PVI) procedures. We evaluated the influence of catheter orientation on lesion dimensions in an in vitro experimental setting and investigated the catheter orientation behavior during PVI.Methods and results: 72 lesions were created in vitro on a porcine heart using a contact force catheter in a parallel, oblique, and perpendicular tip to tissue orientation. The superficial lesion length (SLL) increased shifting the catheter from perpendicular to parallel orientation. The intratissue absolute maximal lesion length (AML) was greater with an oblique catheter orientation. The lesion depth (LD) and the superficial lesion width (SLW) resulted similar with any orientation.Data from 21 PVI procedures in patients with paroxysmal atrial fibrillation using the “wide antral circumferential RF ablation” (WACA) technique were retrospectively analyzed. The mean contact angle among 1130 RF lesions was 28 ± 20°. A prevalent parallel orientation was noted in the anterior WACA segments, whereas it resulted more perpendicular in the posterior segments. Significant differences in catheter orientation between the three operators were found only in few WACA segments.Conclusions: In an in vitro setting, catheter orientation affects SLL and AML, but not LD and SLW. During PVI procedures, catheter orientation resulted mostly parallel to the endocardium. Catheter orientation varied among different WACA segments, but only small differences were found between three operators when considering similar WACA segments.Keywords
This publication has 21 references indexed in Scilit:
- Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine HeartCirculation: Arrhythmia and Electrophysiology, 2014
- Factors Affecting Catheter Contact in the Human Left Atrium and Their Impact on Ablation EfficacyJournal of Cardiovascular Electrophysiology, 2014
- Contact Geometry Affects Lesion Formation in Radio-Frequency Cardiac Catheter AblationPLOS ONE, 2013
- Ablation Effect Indicated by Impedance Fall is Correlated with Contact Force Level During Ablation for Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2013
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA studyHeart Rhythm, 2012
- Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile ModelJournal of Cardiovascular Electrophysiology, 2010
- Effect of Electrode Orientation on Lesion Sizes Produced by Irrigated Radiofrequency Ablation CathetersJournal of Cardiovascular Electrophysiology, 2009
- Comparison of Lesion Sizes Produced by Cryoablation and Open Irrigation Radiofrequency Ablation CathetersJournal of Cardiovascular Electrophysiology, 2008
- Mechanisms for Enlarging Lesion Size During Irrigated Tip Radiofrequency Ablation:. Is There a Virtual Electrode Effect?Journal of Interventional Cardiology, 2004