Open‐source implementation of X‐nuclear sequences using the Pulseq framework
- 2 April 2025
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 94 (2) , 651-664
Abstract
Purpose: Create vendor‐neutral modular sequences for X‐nuclear acquisitions and build an X‐nuclear–enabled Pulseq interpreter for GE (GE HealthCare, Waukesha, WI) scanners. Methods: We designed a modular 2D gradient echo spiral sequence to support several sequence formats and a modular metabolite‐specific 3D balanced steady‐state free precession sequence for hyperpolarized (HP) carbon‐13 ( 13 C) MRI. In addition, we developed a new Pulseq interpreter for GE scanners, named TOPPE MNS (TOPPE Multi‐Nuclear Spectroscopy), to implement X‐nuclear acquisitions capabilities. We evaluated TOPPE MNS and the modular sequences through phantom studies using phosphorus‐31 ( 31 P), hydrogen‐2 ( 2 H), and 13 C coils, and in vivo studies including a human brain deuterium metabolic imaging study at natural abundance, HP 13 C animal studies, and human renal studies. Results: Data from the 13 C phantom showed the accuracy of designed modular sequences and consistent performance with the product sequences. 31 P, 2 H, and 13 C phantom studies and a multi‐vendor/multi‐version 13 C phantom study showed accurate excitation and spatial encoding functionalities. A 2 H‐MRS brain volunteer study, HP [1‐ 13 C]pyruvate animal study, and human renal study showed good image quality with SNR comparable to those reported in the published literature. These results demonstrated the reproducibility of the TOPPE MNS GE interpreter and modular spiral sequences. Conclusion: We have designed a modular 2D gradient echo spiral sequence supporting several sequence formats and a modular metabolic‐specific 3D balanced steady‐state free precession sequence for 13 C acquisition, as well as developed a GE interpreter with X‐nucleus capabilities. Our work paves the way for future multi‐site studies with acquisitions for X‐nuclei across MRI vendors and software versions.Keywords
Funding Information
- Deutschen Konsortium für Translationale Krebsforschung
- National Institutes of Health (R01CA249099, R21NS133531, P41EB013598, R01AG080981, R01CA280071, S10OD025077)
This publication has 27 references indexed in Scilit:
- A regional bolus tracking and real‐time B1 calibration method for hyperpolarized 13C MRIMagnetic Resonance in Medicine, 2018
- TOPPE: A framework for rapid prototyping of MR pulse sequencesMagnetic Resonance in Medicine, 2017
- Pulseq: A rapid and hardware-independent pulse sequence prototyping frameworkMagnetic Resonance in Medicine, 2016
- Development of a symmetric echo planar imaging framework for clinical translation of rapid dynamic hyperpolarized 13C imagingMagnetic Resonance in Medicine, 2016
- ESPIRiT—an eigenvalue approach to autocalibrating parallel MRI: Where SENSE meets GRAPPAMagnetic Resonance in Medicine, 2013
- IDEAL spiral CSI for dynamic metabolic MR imaging of hyperpolarized [1‐13C]pyruvateMagnetic Resonance in Medicine, 2011
- Two‐dimensional phase cycled reconstruction for inherent correction of echo‐planar imaging nyquist artifactsMagnetic Resonance in Medicine, 2011
- Application of subsecond spiral chemical shift imaging to real‐time multislice metabolic imaging of the rat in vivo after injection of hyperpolarized 13C1‐pyruvateMagnetic Resonance in Medicine, 2009
- SENSE: sensitivity encoding for fast MRI.1999
- Sampling density compensation in MRI: Rationale and an iterative numerical solutionMagnetic Resonance in Medicine, 1999