Dosimetric Study of Flattened Versus Unflattened Filter-Free Medical Linear Accelerator: Experimental Measurements and Monte Carlo Calculations
Open Access
- 24 February 2025
- journal article
- research article
- Published by MDPI AG in Quantum Beam Science
- Vol. 9 (1) , 7
Abstract
A dosimetric study compared flattened filter (FF) and unflattened filter-free (FFF) 18 MV medical linear accelerators (LINAC) using BEAMnrc Monte Carlo (MC) calculations and experimental measurements. BEAMnrc MC simulations were initially validated against experimental measurements for an 18 MV FF LINAC, with parameters such as the percentage depth dose (PDD) and beam profile measured and calculated per the International Atomic Energy Agency (IAEA) dosimetry protocol TRS 398. Following the validation of the LINAC and water phantom models for MC simulations, BEAMnrc MC calculations were performed to compare the FF and FFF 18 MV LINAC parameters. The results indicate that the BEAMnrc MC accurately simulated the LINAC model, with PDD uncertainties within 2%. Beam flatness differences between the MC simulations and measurements in the plateau region were within 3% and within 2 mm in the penumbra region. The PDD data show that the 18 MV FFF beam delivered a higher dose rate in the buildup region than the FF beam, while beam profile measurements indicate lower out-of-field doses for FFF beams, especially in the 20 × 20 cm2 field. These findings provide crucial dosimetric data for an 18 MV FFF LINAC, which is useful for quality assurance and beam matching, and offer a methodology for quantitatively comparing the dosimetry properties of an individual 18 MV FFF LINAC to reference data.Keywords
Funding Information
- Deanship of Scientific Research at Imam Mohammad Ibn Saud Islamic University (IMSIU) (IMSIU-RPP2023079)
This publication has 14 references indexed in Scilit:
- Analysis of Uncertainties in Clinical High-Energy Photon Beam Calibrations Using Absorbed Dose StandardsApplied Sciences, 2022
- Comparison of Dosimetry Protocols for Electron Beam Radiotherapy Calibrations and Measurement UncertaintiesLife, 2021
- Progress and prospects of flattening filter free beam technology in radiosurgery and stereotactic body radiotherapyCritical Reviews in Oncology/Hematology, 2021
- Beam matching of two linear accelerators, identical model and brandJournal of Physics: Conference Series, 2020
- Dosimetric properties of equivalent-quality flattening filter-free (FFF) and flattened photon beams of Versa HD linear acceleratorJournal of Applied Clinical Medical Physics, 2016
- Varian 2100C/D Clinac 18 MV photon phase space file characterization and modeling by using MCNP CodeThe European Physical Journal Plus, 2015
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: development and validationPhysics in Medicine & Biology, 2009
- Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear acceleratorRadiation Medicine, 2008
- Efficient photon beam dose calculations using DOSXYZnrc with BEAMnrcMedical Physics, 2006