Abstract
Pressure sensitive mats (PSMs) have a wide variety of applications. They can be used to measure the center of pressure (CoP) when a person is standing. Measuring CoP is useful for several applications, such as human stability assessment or fall risk prediction. Low-cost PSMs can be manufactured using piezoresistive materials, such as Velostat®, whose conductance depends on the pressure exerted. However, these low-cost materials often suffer from hysteresis and creep. These effects worsen CoP measurement. Therefore, correction of hysteresis and creep effects is required to improve it. In this study, a creep-hysteresis correction model is applied to quantify CoP displacements obtained with PSMs. The model to correct creep and hysteresis effects is optimized using the CoP measurements themselves. This is a novel approach to optimize the creep and hysteresis model for this application. This approach has been experimentally tested on three different PSMs, including a commercial mat. Several tests and comparisons with existing models were performed, resulting in an improvement in CoP measurement when the CoP-optimized creep-hysteresis correction model was applied.
Funding Information
  • Ministerio de Ciencia, Innovaci?n y Universidades (EST24/00517, FPU20/04578, PID2021-125091OB-I00)
  • Gobierno de Arag?n (T49_23)
  • Fundaci?n Bancaria Ibercaja (IT 1/24)

This publication has 42 references indexed in Scilit: