Influence of acetabular cup thickness on seating and primary stability in total hip arthroplasty
Authors:
Open Access
- 2 December 2021
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 40 (9) , 2139-2146
Abstract
Insufficient primary stability of acetabular hip cups is a complication resulting in early cup loosening. Available cup designs vary in terms of wall thickness, potentially affecting implant fixation. This study investigated the influence of different wall thicknesses on the implantation process and the resulting primary stability using excised human acetabula. Implantations were performed using a powered impaction device providing consistent energy with each stroke. Two different wall thicknesses were compared in terms of seating progress, polar gap remaining after implantation, bone‐to‐implant contact area, cup deflection, and lever out moment. Thin‐walled cups showed higher lever out resistance (p < 0.001) and smaller polar gaps (p < 0.001) with larger bone contact toward the dome of the cup (p < 0.001) compared to thick‐walled cups. Small seating steps at the end of the impaction process were observed if a high number of strokes were needed to seat the cup (p = 0.045). A high number of strokes led to a strain release of the cup during the final strokes (p = 0.003). This strain release is indicative for over‐impaction of the cup associated with bone damage and reduced primary stability. Adequate cup seating can be achieved with thin‐walled cups with lower energy input in comparison to thicker ones. Thin‐walled cups showed improved primary stability and enable implantation with lower energy input, reducing the risk of over‐impaction and bone damage. Additional strokes should be avoided as soon as no further seating progress has been observed.Keywords
This publication has 35 references indexed in Scilit:
- Early proximal migration of cups is associated with late revision in THAActa Orthopaedica, 2012
- The influence of acetabular cup material on pelvis cortex surface strains, measured using digital image correlationJournal of Biomechanics, 2012
- Deformation characteristics and eigenfrequencies of press-fit acetabular cupsClinical Biomechanics, 2011
- The Effects of Acetabular Shell Deformation and Liner Thickness on Frictional Torque in Ultrahigh-Molecular-Weight Polyethylene Acetabular BearingsThe Journal of Arthroplasty, 2010
- The Effect of Under-Reaming on the Cup/Bone Interface of a Press Fit Hip ReplacementJournal of Biomechanical Engineering, 2010
- Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press‐fit implantationJournal of Orthopaedic Research, 2009
- Monitoring Teriparatide-Associated Changes in Vertebral Microstructure by High-Resolution CT In Vivo: Results From the EUROFORS StudyJournal of Bone and Mineral Research, 2007
- The effect of interfacial parameters on cup–bone relative micromotionsJournal of Biomechanics, 2001
- The Harris-Galante porous acetabular component press-fit without screw fixationThe Journal of Arthroplasty, 1994
- Stability of press-fit acetabular cupsThe Journal of Arthroplasty, 1992