Physical Characteristics of Polyaxial-Headed Pedicle Screws and Biomechanical Comparison of Load With Their Failure
Authors:
- Guy R. FogelGFGuy R. Fogel
- Charles A. ReitmanCRCharles A. Reitman
- Weiqiang LiuWLWeiqiang Liu
- Stephen I. EssesSEStephen I. Esses
- 1 March 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 28 (5) , 470-473
Abstract
Pedicle screw strength or load to failure was biomechanically evaluated, and the geometric characteristics of pedicle screw instrumentation systems were compared. To compare the features of pedicle screw systems, and to demonstrate the failure point of the polyaxial pedicle screw head. Many pedicle screw instrumentation systems are currently available to the spine surgeon. Each system has its unique characteristics. It is important for the surgeon to understand the differences in these pedicle screw systems. Pedicle screw load to failure has not been subjected to a comparison study. The physical characteristics of each pedicle screw instrumentation system were determined. Features of rods, instruments, and pedicle screws were cataloged. Biomechanical testing of the pedicle screw construct was performed to determine the site and force of the load to failure. Nine pedicle screw systems were evaluated. Testing was performed with a pneumatic testing system under load control. Three polyaxial screws were used for each test at a load rate of 100 N/second. The load failure value was the force at which the pedicle screw or polyaxial head-screw interface initially deflected. Biomechanical testing demonstrated in all instances that the polyaxial head coupling to the screw was the first failure point. Although there have been subtle design differences in the instruments over time, the features of the pedicle screw instrument sets have become remarkably similar. Biomechanical pedicle screw load-to-failure data demonstrated that the polyaxial head coupling to the screw is the first to fail and may be a protective feature of the pedicle screw, preventing pedicle screw breakage. Knowing the physical characteristics of the available pedicle screw instrumentation systems may allow the choice of pedicle screw best suited for a given clinical situation.Keywords
This publication has 12 references indexed in Scilit:
- Improving the Pullout Strength of Pedicle Screws by Screw CouplingJournal of Spinal Disorders, 2001
- Triangulated Pedicle Screw Construct Technique and Pull-Out Strength of Conical and Cylindrical ScrewsJournal of Spinal Disorders, 2001
- Loosening of sacral screw fixation under in vitro fatigue loadingJournal of Orthopaedic Research, 2000
- Comparison of the In Vitro Holding Strengths of Conical and Cylindrical Pedicle Screws in a Fully Inserted Setting and Backed Out 180°Journal of Spinal Disorders, 2000
- Biomechanical Study of Lumbar Pedicle ScrewsJournal of Spinal Disorders, 1998
- Biomechanics of Transfixation in Pedicle Screw InstrumentationSpine, 1996
- Biomechanical Properties of Threaded Inserts for Lumbar Interbody Spinal FusionSpine, 1995
- Pedicle-Screw Fixation in the Lumbar SpineJournal of the American Academy of Orthopaedic Surgeons, 1995
- The Load Sharing Classification of Spine FracturesSpine, 1994
- Correlations Between Screw Hole Preparation, Torque of Insertion, and Pullout Strength for Spinal ScrewsJournal of Spinal Disorders, 1994