Wavelength Dispersive X-ray Fluorescence Imaging

Authors:

Abstract
A new wavelength-dispersive X-ray fluorescence (WD-XRF) imaging spectrometer equipped with a two-dimensional X-ray detector was developed in the laboratory. Straight polycapillary optics was applied instead of a soller slit, which is used in conventional WD-XRF spectrometers. X-rays were guided through the straight polycapillary to the exit of the optics by X-ray external total reflections. X-ray fluorescence was dispersed by an analyzing crystal (LiF(200)), keeping the information of elemental distribution on the surface of the sample. The energy resolution of the developed spectrometer was 130–152 eV at the Zn Kα peak. X-ray elemental images of Cu Kα and Ni Kα were successfully obtained by an X-ray CCD detector at the corresponding diffraction angles. The analytical performance of this technique, and further improvements are discussed.