Investigation into dosimetric effect of a MAGNA‐SITE™ tissue expander on post‐mastectomy radiotherapy
Authors:
- Rosemary C. A. ThompsonRTRosemary C. A. Thompson
- Andrew M. MorganAMAndrew M. Morgan
- 20 May 2005
- journal article
- Published by Wiley in Medical Physics
- Vol. 32 (6Part1) , 1640-1646
Abstract
It is increasingly common for radiotherapy departments to encounter high density objects in patients being planned for radiotherapy. Many cases, such as artificial hip prostheses, are well documented. In our cancer center we have recently come across a new type of implant—a McGhan Style 133 Tissue Expander—implanted in patients who have undergone a mastectomy and will in future have breast reconstruction. This type of implant contains a small rare earth magnet encased in a titanium body, which induces significant perturbations in the dose distribution. These perturbations have been measured using a p‐type semiconductor diode. Attenuation of up to 30% of local dose has been observed for a single beam. However, in more realistic clinical situations using tangential parallel opposed beams, it is estimated that part of the planning target volume maybe be underdosed by approximately 10%. Comparisons have been made between measured attenuation and that calculated by a treatment planning system, which demonstrates inadequate modeling of the dose perturbation caused by the implant in this case.Keywords
This publication has 11 references indexed in Scilit:
- Dosimetric verification of a commercial collapsed cone algorithm in simulated clinical situationsRadiotherapy and Oncology, 2004
- Experimental verification of convolution/superposition photon dose calculations for radiotherapy treatment planningPhysics in Medicine & Biology, 2003
- Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63Medical Physics, 2003
- How accurate is a CT-based dose calculation on a pencil beam TPS for a patient with a metallic prosthesis?Physics in Medicine & Biology, 2001
- Evaluation of the dosimetric characteristics of a diamond detector for photon beam measurementsMedical Physics, 1995
- A Petrov-Galerkin spectral element technique for heterogeneous porous media flowComputers & Mathematics with Applications, 1995
- A pencil beam model for photon dose calculationMedical Physics, 1992
- Radiotherapy and breast reconstruction: Clinical results and dosimetryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991
- Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous mediaMedical Physics, 1989
- Penetration of backscattered electrons in polystyrene for energies between 1 and 25 MeVPhysics in Medicine & Biology, 1982