Determination of background, signal‐to‐noise, and dynamic range of a flow cytometer: A novel practical method for instrument characterization and standardization
- 27 September 2017
- journal article
- research article
- Published by Wiley in Cytometry Part A
- Vol. 91 (11) , 1104-1114
Abstract
A well‐defined scale calibration in flow cytometry can improve many aspects of data acquisition such as cytometer setup, instrument comparison and sample comparison. The theory for scale calibration was proposed by Steen over two decades ago, but it has never been put into regular use due to the lack of a widely available precision light source. The introduction of such a light source, the quantiFlashTM , gave this possibility. Here, we describe how this light source can be used to characterize a cytometer's PMT performance. We, therefore, characterized the instrument's response over the entire PMT voltage range. As a consequence, we propose a practical method to characterize a cytometer's signal‐to‐noise ratio (SNR) and dynamic range (DNR). This allows the selection of a voltage/gain corresponding to a PMT's maximum efficiency and hence the lowest electronic noise, which can help with experiment design. We further introduced a decibel (dB) scale for the presentation of SNR and DNR values. SNR and DNR are stand‐alone values that allow the direct comparison of different instruments. Finally, with this method, it becomes clear that increased SNR comes at the expense of DNR and thus the limiting factor of modern cytometers is the DNR. © 2017 International Society for Advancement of CytometryKeywords
Funding Information
- Berlin-Brandenburg School for Regenerative Therapies (GSC 203)
- European Regional Development Fund (Europäischer Fonds für regionale Entwicklung (EFRE)), Germany (10158448/9)
This publication has 15 references indexed in Scilit:
- Evaluating flow cytometer performance with weighted quadratic least squares analysis of LED and multi‐level bead dataCytometry Part A, 2017
- Standardization, Calibration, and Control in Flow CytometryCurrent Protocols in Cytometry, 2017
- Q and B values are critical measurements required for inter‐instrument standardization and development of multicolor flow cytometry staining panelsCytometry Part A, 2014
- Quality assurance for polychromatic flow cytometry using a suite of calibration beadsNature Protocols, 2012
- EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocolsLeukemia, 2012
- An LED pulser for measuring photomultiplier linearityNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012
- Methods for Accurate Measurement of the Response of Photomultiplier Tubes and Intensity of Light PulsesIEEE Transactions on Nuclear Science, 2011
- Flow cytometry controls, instrument setup, and the determination of positivityCytometry Part A, 2006
- Evaluating fluorescence sensitivity on flow cytometers: An overviewCytometry, 1998
- Noise, Sensitivity, and Resolution of Flow CytometersCytometry, 1992