Energy-Resolved CNR Performance in Dense-Breast and Implant X-Ray Mammography Using a CdTe Photon-Counting Detector: A Monte Carlo Study
Authors:
- Gerardo RoqueGRGerardo Roque
- Maria Laura Pérez-LaraMPMaria Laura Pérez-Lara
- Steven CelySCSteven Cely
- Juan Sebastián Useche ParraJPJuan Sebastián Useche Parra
- Jesús David BermúdezJBJesús David Bermúdez
- Michael K. SchützMSMichael K. Schütz
- Michael FiederleMFMichael Fiederle
- Carlos ÁvilaCÁCarlos Ávila
- Simon ProczSPSimon Procz
Open Access
- 5 April 2026
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 16 (7) , 3550
Abstract
X-ray imaging of dense breasts and breast implants often suffers from reduced lesion visibility because strong attenuation lowers contrast, while conventional rhodium (Rh) K-edge filtering suppresses part of the high-energy spectral tail. This study presents a Monte Carlo framework for spectroscopic mammography using a voxelated 1 mm thick cadmium telluride (CdTe) sensor and a first-order detector interaction model to evaluate energy-dependent image quality. The model reproduces fluorescence and inter-voxel energy redistribution in CdTe, but not the full detector chain, and remains idealized with respect to charge transport, carrier collection, threshold dispersion, and pile-up. Energy-resolved simulations in the 10–50 keV range were used to compute spectroscopic contrast-to-noise ratio (CNR) curves and to form integrated-spectrum (IS) images for four tested spectra. For the dense-breast calcium hydroxyapatite (HA) speck detection task considered here, and under the present simulation assumptions, replacing the standard 28 kVp + 50 μm Rh spectrum with 28 kVp + 1 mm Al increased the simulated IS image CNR by 23.11%, with an approximately 5% increase in estimated primary-incident air kerma at the phantom entrance plane. Preliminary experimental implant-phantom images were included as a qualitative feasibility check, showing a trend consistent with simulations. Within the limits of this task-specific simulation, the results suggest that preserving the transmitted high-energy tail can improve HA speck visibility for the present 1 mm CdTe photon-counting detector, with the 28 kVp + 1 mm Al spectrum outperforming the other tested cases.Keywords
All Related Versions
Funding Information
- Bundesministerium für Forschung, Technologie und Raumfahrt (01DN21005)
- Ministerio de Ciencia, Tecnología e Innovación (873-2020)
- Universidad de Los Andes, Faculty of Science (INV-2025-213-3462)
This publication has 51 references indexed in Scilit:
- Technical note: SpekPy Web—online x‐ray spectrum calculations using an interface to the SpekPy toolkitJournal of Applied Clinical Medical Physics, 2024
- Technical Note: SpekPy v2.0—a software toolkit for modeling x‐ray tube spectraMedical Physics, 2021
- RECORDS: improved Reporting of montE CarlO RaDiation transport Studies: Report of the AAPM Research Committee Task Group 268Medical Physics, 2017
- Investigating the feasibility of classifying breast microcalcifications using photon-counting spectral mammography: A simulation studyMedical Physics, 2017
- BI-RADS ® fifth edition: A summary of changesDiagnostic and Interventional Imaging, 2017
- Screening ultrasound as an adjunct to mammography in women with mammographically dense breastsAmerican Journal of Obstetrics and Gynecology, 2015
- The influence of energy weighting on X-ray imaging qualityNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- Effect of Breast Augmentation on the Accuracy of Mammography and Cancer CharacteristicsJAMA, 2004
- Mammography and breast implantsBritish Journal of Plastic Surgery, 1995
- Vision1973