Lead Thrombi Associated with Externalized Cables on Riata ICD Leads: A Case Series
- 11 April 2013
- journal article
- case report
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 24 (9) , 1047-1050
Abstract
Thrombi on Riata ICD Leads: St. Jude Riata family ICD leads are subject to an FDA class I recall due to insulation failure, cable extrusions, and a high rate of electrical failures. We present multiple cases of large intravascular thrombus formation adherent to externalized conductor cables on Riata leads. Our observations highlight a previously unknown risk of large or calcified thrombus formation involving externalized conductors. We suggest that these leads should be screened for thrombi prior to potential lead extraction to determine the safest approach. Serial echocardiography may help detect progressive thrombus formation. The role of systemic anticoagulation for prevention or treatment of these thrombi remains unclear.Keywords
This publication has 12 references indexed in Scilit:
- Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leadsHeart Rhythm, 2013
- Extinction of small-caliber transvenous ICD leads: Downsizing in a race to a recallHeart Rhythm, 2013
- Failure of implantable cardioverter-defibrillator leads: A matter of lead size?Heart Rhythm, 2013
- Failure of a novel silicone-polyurethane copolymer (Optim ) to prevent implantable cardioverter-defibrillator lead insulation abrasionsEP Europace, 2012
- Class I recall of defibrillator leads: A comparison of the Sprint Fidelis and Riata familiesHeart Rhythm, 2012
- High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leadsHeart Rhythm, 2012
- Prevalence and Predictors of Cable Extrusion and Loss of Electrical Integrity with the Riata Defibrillator LeadJournal of Cardiovascular Electrophysiology, 2012
- Failure to Deliver Therapy by a Riata Lead with Internal Wire Externalization and Normal Electrical Parameters During Routine InterrogationJournal of Cardiovascular Electrophysiology, 2012
- Insulation Defects of Thin High-Voltage ICD Leads:Journal of Cardiovascular Electrophysiology, 2011
- Premature Failure of a Riata Defibrillator Lead Without Impedance Change or Inappropriate Sensing: A Case Report and Review of the LiteratureJournal of Cardiovascular Electrophysiology, 2011