Unrestrained acoustic plethysmograph for measuring specific airway resistance in mice
Authors:
- Jeffrey S. ReynoldsJRJeffrey S. Reynolds
- Victor J. JohnsonVJVictor J. Johnson
Open Access
- 1 August 2008
- journal article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 105 (2) , 711-717
Abstract
An acoustic whole body plethysmograph was developed to estimate specific airway resistance (sRaw) in unrestrained mice. The plethysmograph uses acoustic principles to measure the thoracic breathing pattern and simultaneously measures the airflow entering and/or leaving the plethysmograph. Similarly to traditional methods utilizing a double-chamber plethysmograph, these measurements were combined to estimate sRaw. To evaluate the new system, we placed six conscious A/J mice individually in a whole body plethysmograph (Buxco System) for a 2-min exposure to aerosolized methacholine chloride dissolved in saline (0, 5, 10, and 20 mg/ml), which is known to increase sRaw in mice. Three minutes after exposure, the mice were transferred to the acoustic plethysmograph for 2 min for data collection. The mean baseline value of sRaw was 0.93 ± 0.10 cmH2O·s. A dose-dependent increase in sRaw was shown, with an approximate tripling of sRaw at the highest dose. These results demonstrate the ability of the system to estimate sRaw based on plethysmograph airflow and acoustic amplitude.Keywords
This publication has 23 references indexed in Scilit:
- Unrestrained video-assisted plethysmography: a noninvasive method for assessment of lung mechanical function in small animalsJournal of Applied Physiology, 2008
- Restrained whole body plethysmography for measure of strain-specific and allergen-induced airway responsiveness in conscious miceJournal of Applied Physiology, 2006
- Unrestrained Acoustic Plethysmograph for Measuring Tidal Volume in MiceAnnals of Biomedical Engineering, 2006
- Further exploration of the Penh parameterExperimental and Toxicologic Pathology, 2006
- Unrestrained plethysmography is an unreliable measure of airway responsiveness in BALB/c and C57BL/6 miceJournal of Applied Physiology, 2004
- MOUSE MODELS OF AIRWAY RESPONSIVENESS: Physiological Basis of Observed Outcomes and Analysis of Selected Examples Using These Outcome IndicatorsAnnual Review of Physiology, 1999
- Whole-body plethysmography, does it measure tidal volume of small animals?Canadian Journal of Physiology and Pharmacology, 1998
- Noninvasive Measurement of Airway Responsiveness in Allergic Mice Using Barometric PlethysmographyAmerican Journal of Respiratory and Critical Care Medicine, 1997
- A theoretical analysis of the barometric method for measurement of tidal volumeRespiration Physiology, 1978
- Temperature gradients in the nasal passage of some small mammalsComparative Biochemistry and Physiology Part A: Physiology, 1976