In Vitro Reproduction of Device Deformation Leading to Thrombotic Complications and Failure of Flow Diversion
- 1 December 2013
- journal article
- research article
- Published by SAGE Publications in Interventional Neuroradiology
- Vol. 19 (4) , 432-437
Abstract
Flow diverters (FDs) are increasingly used for complex intracranial aneurysms. As these self-expanding devices are deployed across an aneurysm neck, they can undergo deformations. The potential clinical consequences of FD deformations remain unclear. We describe an immediate thrombotic complication attributed to a stereotypical stenotic deformation of an FD extremity that can occur when landing zones are of insufficient length. This case is supplemented with in vitro studies showing the relationship between i) the length of the landing zones and ii) discrepancies between the diameter of the device and recipient vessel, and the severity of FD stenosis. In vitro, a shorter landing zone was associated with a progressive stenotic deformation of the terminal ends of all FDs studied. This deformation was more pronounced when the diameter of the device was oversized compared to the size of the recipient tube. In our clinical case, the presence of this deformation led to an immediate thrombotic complication, requiring deployment of a second stent to correct the observed stenosis. In addition, treatment failure ultimately led to a fatal rupture, a failure that can be explained by residual flows through a more porous transition zone, another characteristic FD deformation which occurs when they are oversized as compared to the parent vessel, but free to expand at the level of the aneurysm. Proper selection of device diameter and length of the landing zone is important, and may decrease the incidence of deformation-related complications.Keywords
This publication has 14 references indexed in Scilit:
- Stents and flow diverters in the treatment of aneurysms: device deformation in vivo may alter porosity and impact efficacyNeuroradiology, 2012
- The Varying Porosity of Braided Self-Expanding Stents and Flow Diverters: An Experimental StudyAmerican Journal of Neuroradiology, 2012
- Immediate and Midterm Results following Treatment of Unruptured Intracranial Aneurysms with the Pipeline Embolization DeviceAmerican Journal of Neuroradiology, 2011
- The Silk Flow-Diverting Stent in the Endovascular Treatment of Complex Intracranial Aneurysms: Technical Aspects and Midterm Results in 24 Consecutive PatientsNeurosurgery, 2011
- Rescue, Combined, and Stand-Alone Thrombectomy in the Management of Large Vessel Occlusion Stroke Using the Solitaire Device: A Prospective 50-Patient Single-Center StudyStroke, 2011
- Intra-Aneurysmal Thrombosis as a Possible Cause of Delayed Aneurysm Rupture after Flow-Diversion TreatmentAmerican Journal of Neuroradiology, 2010
- Treatment of Intracranial Aneurysms by Functional Reconstruction of the Parent Artery: The Budapest Experience with the Pipeline Embolization DeviceAmerican Journal of Neuroradiology, 2010
- CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- DEFINITIVE RECONSTRUCTION OF CIRCUMFERENTIAL, FUSIFORM INTRACRANIAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2008
- Mechanisms of Occlusion and Recanalization in Canine Carotid Bifurcation Aneurysms Embolized with Platinum Coils: An Alternative ConceptAmerican Journal of Neuroradiology, 2008