Fundamental Evaluation of Segmented Attenuation Correction Method for Clinical FDG-PET Studies : Simulation of Pulmonary Mass Lesions in Phantom Studies
Authors:
Open Access
- 1 January 2002
- journal article
- Published by Japanese Society of Radiological Technology in Japanese Journal of Radiological Technology
- Vol. 58 (12) , 1579-1585
Abstract
Objective: Both the segmented attenuation correction (SAC) method and post-injection transmission scanning are useful and widespread in clinical whole-body FDG-PET studies. The SAC method usually accomplishes smoothing of the transmission data. This calculation segments a μ-map into three degrees (lung, soft tissue, and bone) of attenuation coefficient. This method is used to reduce transmission scan time without deteriorating the quality of PET images. However, the SAC method has a tendency to underestimate the attenuation coefficient, resulting lower detectability for lung field mass lesions. We therefore evaluated the quantitative accuracy of the SAC method using transmission scanning and emission scanning data in a phantom study. Methods: A dedicated 3D PET scanner, the Siemens ECAT EXACT HR+, was used to scan images of two types of phantoms, a spherical phantom (Japan Radioisotope Association phantom) and a cylindrical phantom (20 cm in diameter). We evaluated differences between transmission images (μ-map) of the SAC method and measured attenuation correction (MAC)method, these two kinds of attenuation-corrected emission data (emission + SAC method, emission + MAC method), and emission data only (without attenuation correction). Results: In the μ-map, recovery coefficient (RC) values at 10 mm in diameter were 0.27 and 0.00 in the MAC and SAC methods, respectively, in the spherical hot area. For the emission data, the emission + SAC method and emission + MAC method showed almost the same RC values for all sizes of hot area diameter. The SAC method, however, resulted in 20% underestimation for all sizes of hot area diameter as compared with the MAC method. Conclusion: In pulmonary mass lesions, it is necessary to correct for the partial volume effect in quantitative PET measurement. However, from our data, the SAC method is not appropriate for partial volume effect correction.Keywords
This publication has 7 references indexed in Scilit:
- Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- FDG-PET standardized uptake values in normal anatomical structures using iterative reconstruction segmented attenuation correction and filtered back-projectionEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- An automatic classification technique for attenuation correction in positron emission tomographyEuropean Journal of Nuclear Medicine and Molecular Imaging, 1999
- Adaptive, segmented attenuation correction for whole-body PET imagingIEEE Transactions on Nuclear Science, 1996
- Investigation of partial volume correction methods for brain FDG PET studiesIEEE Transactions on Nuclear Science, 1996
- Local threshold for segmented attenuation correction of PET imaging of the thoraxIEEE Transactions on Nuclear Science, 1994
- Analysis of Emission Tomographic Scan Data: Limitations Imposed by Resolution and BackgroundJournal of Computer Assisted Tomography, 1984