The sterilization efficacy of reprocessed single use diathermy pencils.
Open Access
- 1 February 2010
- journal article
- research article
- Published by FapUNIFESP (SciELO) in Revista Latino-Americana de Enfermagem
- Vol. 18 (1) , 81-86
Abstract
In Brazil, single use diathermy pencils (SUDP) are among the most common reused devices. This study assesses the sterilization efficacy of reprocessing SUDP using two cleansing methods (manual or automated), followed by one of three of the low-temperature sterilization methods: Hydrogen Peroxide Plasma (HPP), Ethylene Oxide (ETO) or Low-Temperature Steam Formaldehyde (LTSF). The sample was composed of 360 SUDP after their first use. The probability of sterilization failure was estimated considering the number of positive microbiological results obtained by cultures of the studied devices. The overall sterilization failure probability for SUDP was 0.26. The sterilization method, which presented the lowest failure probability was the LTSF (0.01), followed by ETO (0.21) and HPP (0.56). Automated cleansing obtained a better result than manual cleansing. This trial demonstrated that the probability of sterilization in reprocessed SUDP is highly dependent on both the type of cleansing and the sterilization method applied.Keywords
This publication has 11 references indexed in Scilit:
- Reprocessing and Reuse of Single-Use Medical Devices Used During Hemodynamic Procedures in Brazil: A Widespread and Largely Overlooked ProblemInfection Control & Hospital Epidemiology, 2008
- Low-temperature sterilization and new technologies.Revista Latino-Americana de Enfermagem, 2007
- Análise dos custos do reprocessamento de pinças de uso único utilizadas em cirurgia vídeo-assistidaRevista Latino-Americana de Enfermagem, 2006
- Critérios para avaliação das dificuldades na limpeza de artigos de uso únicoRevista Latino-Americana de Enfermagem, 2006
- A Comparative Study of Ethylene Oxide Gas, Hydrogen Peroxide Gas Plasma, and Low-Temperature Steam Formaldehyde SterilizationInfection Control & Hospital Epidemiology, 2005
- Manual versus automated methods for cleaning reusable accessory devices used for minimally invasive surgical proceduresJournal of Hospital Infection, 2004
- Inadequacy of manual cleaning for reprocessing single-use, triple-lumen sphinctertomes: Simulated-use testing comparing manual with automated cleaning methodsAmerican Journal of Infection Control, 2003
- Can Labeled Single-Use Devices Be Reused? An Old Question in the New EraJournal of Vascular and Interventional Radiology, 2002
- Specific hygiene issues relating to reprocessing and reuse of single-use devices for laparoscopic surgerySurgical Endoscopy, 2002
- Comparison of Ion Plasma, Vaporized Hydrogen Peroxide, and 100% Ethylene Oxide Sterilizers to the 12/88 Ethylene Oxide Gas SterilizerInfection Control & Hospital Epidemiology, 1996