Impact of Alternate b-Value Combinations and Metrics on the Predictive Performance and Repeatability of Diffusion-Weighted MRI in Breast Cancer Treatment: Results from the ECOG-ACRIN A6698 Trial
Authors:
- Savannah C. PartridgeSPSavannah C. Partridge
- Jon SteingrimssonJSJon Steingrimsson
- David C. NewittDNDavid C. Newitt
- Jessica E. GibbsJGJessica E. Gibbs
- Helga S. MarquesHMHelga S. Marques
- Patrick J. BolanPBPatrick J. Bolan
- Michael A. BossMBMichael A. Boss
- Thomas L. ChenevertTCThomas L. Chenevert
- Mark A. RosenMRMark A. Rosen
- Nola M. HyltonNHNola M. Hylton
Open Access
- 4 March 2022
- journal article
- research article
- Published by MDPI AG in Tomography
- Vol. 8 (2) , 701-717
Abstract
In diffusion-weighted MRI (DW-MRI), choice of b-value influences apparent diffusion coefficient (ADC) values by probing different aspects of the tissue microenvironment. As a secondary analysis of the multicenter ECOG-ACRIN A6698 trial, the purpose of this study was to investigate the impact of alternate b-value combinations on the performance and repeatability of tumor ADC as a predictive marker of breast cancer treatment response. The final analysis included 210 women who underwent standardized 4-b-value DW-MRI (b = 0/100/600/800 s/mm2) at multiple timepoints during neoadjuvant chemotherapy treatment and a subset (n = 71) who underwent test–retest scans. Centralized tumor ADC and perfusion fraction (fp) measures were performed using variable b-value combinations. Prediction of pathologic complete response (pCR) based on the mid-treatment/12-week percent change in each metric was estimated by area under the receiver operating characteristic curve (AUC). Repeatability was estimated by within-subject coefficient of variation (wCV). Results show that two-b-value ADC calculations provided non-inferior predictive value to four-b-value ADC calculations overall (AUCs = 0.60–0.61 versus AUC = 0.60) and for HR+/HER2− cancers where ADC was most predictive (AUCs = 0.75–0.78 versus AUC = 0.76), p < 0.05. Using two b-values (0/600 or 0/800 s/mm2) did not reduce ADC repeatability over the four-b-value calculation (wCVs = 4.9–5.2% versus 5.4%). The alternate metrics ADCfast (b ≤ 100 s/mm2), ADCslow (b ≥ 100 s/mm2), and fp did not improve predictive performance (AUCs = 0.54–0.60, p = 0.08–0.81), and ADCfast and fp demonstrated the lowest repeatability (wCVs = 6.71% and 12.4%, respectively). In conclusion, breast tumor ADC calculated using a simple two-b-value approach can provide comparable predictive value and repeatability to full four-b-value measurements as a marker of treatment response.Keywords
Funding Information
- National Institutes of Health (U01CA079778, U01CA080098, U10CA180820 and U10CA180794 (ECOG-ACRIN Cancer Research Group), U01CA225427 (N.M.H.), R01CA132870 (N.M.H and D.C.N), R01CA190299 (T.L.C.), R01CA207290 (S.C.P.) and R01CA248192 (S.C.P.))
This publication has 34 references indexed in Scilit:
- Effect of b value and pre-admission of contrast on diagnostic accuracy of 1.5-T breast DWI: a systematic review and meta-analysisEuropean Radiology, 2014
- Diffusion-weighted MRI in pretreatment prediction of response to neoadjuvant chemotherapy in patients with breast cancerEuropean Radiology, 2013
- DW-MRI ADC values can predict treatment response in patients with locally advanced breast cancer undergoing neoadjuvant chemotherapyMedical Oncology, 2011
- The diffusion-weighted imaging perfusion fraction f is a potential marker of sorafenib treatment in advanced hepatocellular carcinoma: a pilot studyEuropean Radiology, 2010
- Diffusion-weighted MR for Differentiation of Breast Lesions at 3.0 T: How Does Selection of Diffusion Protocols Affect Diagnosis?Radiology, 2009
- Assessment of Breast Lesions With Diffusion-Weighted MRI: Comparing the Use of Different b ValuesAmerican Journal of Roentgenology, 2009
- Longitudinal study of the assessment by MRI and diffusion‐weighted imaging of tumor response in patients with locally advanced breast cancer undergoing neoadjuvant chemotherapyNMR in Biomedicine, 2008
- Measurement of Residual Breast Cancer Burden to Predict Survival After Neoadjuvant ChemotherapyJournal of Clinical Oncology, 2007
- Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging.Radiology, 1988
- Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field GradientThe Journal of Chemical Physics, 1965