Evaluation of automated malaria diagnosis using the Sysmex XN-30 analyser in a clinical setting
Open Access
- 22 January 2019
- journal article
- research article
- Published by Springer Nature in Malaria Journal
- Vol. 18 (1) , 1-14
Abstract
Early and accurate diagnosis of malaria is a critical aspect of efforts to control the disease, and several diagnostic tools are available. Microscopic assessment of a peripheral blood smear enables direct visualization of parasites in infected red blood cells and is the clinical diagnostic gold standard. However, it is subjective and requires a high level of skill. Numerous indirect detection methods are in use, but are not ideal since surrogate markers of infection are measured. This study describes the first clinical performance evaluation of the automated Sysmex XN-30 analyser, which utilizes fluorescence flow cytometry to directly detect and quantitate parasite-infected red blood cells. Residual EDTA blood samples from suspected malaria cases referred for routine diagnosis were analysed on the XN-30. Parasitaemia was reported as a percentage, as well as absolute numbers of infected red blood cells, and scattergrams provided a visual image of the parasitized red blood cell clusters. The results reported by the XN-30 correlated with microscopy and the analyser demonstrated 100% sensitivity and specificity. Measurements were reproducible and storage of samples at room temperature did not affect the parameters. Several Plasmodium species were detected, including Plasmodium falciparum, Plasmodium vivax and Plasmodium ovale. The XN-30 also identified the transmissible gametocytes as a separate cluster on the scattergrams. Abnormal red blood cell indices (low haemoglobin and raised reticulocyte counts), haemoglobinopathies and thrombocytopenia did not interfere with the detection of parasites. The XN-30 also generated a concurrent full blood count for each sample. The novel technology of the Sysmex XN-30 provides a robust, rapid, automated and accurate platform for diagnosing malaria in a clinical setting. The objective enumeration of red blood cells infected with Plasmodium species makes it suitable for global use and allows monitoring of the parasite load once therapy has been initiated, thereby providing an early marker of drug resistance. The automated generation of a full blood count for each sample provides an opportunity for detecting unsuspected cases. Asymptomatic carriers can also be identified, which will be useful in blood transfusion centres, and will enable treatment of these individuals to prevent the spread of the disease.Keywords
Funding Information
- Sysmex (Malaria Study)
This publication has 31 references indexed in Scilit:
- What is the best strategy for the prevention of transfusion-transmitted malaria in sub-Saharan African countries where malaria is endemic?Malaria Journal, 2013
- Highly Sensitive Detection of Malaria Parasitemia in a Malaria-Endemic Setting: Performance of a New Loop-Mediated Isothermal Amplification Kit in a Remote Clinic in UgandaThe Journal of Infectious Diseases, 2013
- Human Infections and Detection of Plasmodium knowlesiClinical Microbiology Reviews, 2013
- Haemozoin detection in infected erythrocytes for Plasmodium falciparum malaria diagnosis—Prospects and limitationsActa Tropica, 2012
- Automated detection of malaria with haematology analyzer sysmex xe-2100Indian Journal of Medical Sciences, 2011
- Comparative genomics of the neglected human malaria parasite Plasmodium vivaxNature, 2008
- Development of an Automated Malaria Discriminant Factor Using VCS TechnologyAmerican Journal of Clinical Pathology, 2006
- Development of an Automated Malaria Discriminant Factor Using VCS TechnologyAmerican Journal of Clinical Pathology, 2006
- The Sequence of the Human GenomeScience, 2001
- Screening donor blood for malaria by polymerase chain reactionTransactions of the Royal Society of Tropical Medicine and Hygiene, 1995