Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System
Open Access
- 17 April 2015
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 10 (4) , e0123862
Abstract
The CONCERT-CL closed-loop infusion system designed by VERYARK Technology Co., Ltd. (Guangxi, China) is an innovation using TCI combined with closed-loop controlled intravenous anesthesia under the guide of BIS. In this study we performed a randomized, controlled, multicenter study to compare closed-loop control and open-loop control of propofol by using the CONCERT-CL closed-loop infusion system. 180 surgical patients from three medical centers undergone TCI intravenous anesthesia with propofol and remifentanil were randomly assigned to propofol closed-loop group and propofol opened-loop groups. Primary outcome was global score (GS, GS = (MDAPE+Wobble)/% of time of bispectral index (BIS) 40-60). Secondary outcomes were doses of the anesthetics and emergence time from anesthesia, such as, time to tracheal extubation. There were 89 and 86 patients in the closed-loop and opened-loop groups, respectively. GS in the closed-loop groups (22.21±8.50) were lower than that in the opened-loop group (27.19±15.26) (p=0.009). The higher proportion of time of BIS between 40 and 60 was also observed in the closed-loop group (84.11±9.50%), while that was 79.92±13.17% in the opened-loop group, (p=0.016). No significant differences in propofol dose and time of tracheal extubation were observed. The frequency of propofol regulation in the closed-loop group (31.55±9.46 times/hr) was obverse higher than that in the opened-loop group (6.84±6.21 times/hr) (p=0.000). The CONCERT-CL closed-loop infusion system can automatically regulate the TCI of propofol, maintain the BIS value in an adequate range and reduce the workload of anesthesiologists better than open-loop system. ChiCTR ChiCTR-OOR-14005551Keywords
This publication has 19 references indexed in Scilit:
- Feasibility of Closed-loop Titration of Propofol and Remifentanil Guided by the Spectral M-Entropy MonitorAnesthesiology, 2012
- Closed-Loop Coadministration of Propofol and Remifentanil Guided by Bispectral IndexAnesthesia & Analgesia, 2011
- Comparison of closed loop vs. manual administration of propofol using the Bispectral index in cardiac surgeryActa Anaesthesiologica Scandinavica, 2009
- Closed-Loop Control of Consciousness During Lung Transplantation: An Observational StudyJournal of Cardiothoracic and Vascular Anesthesia, 2008
- Titration of Propofol for Anesthetic Induction and Maintenance Guided by the Bispectral Index: Closed-loop versus Manual ControlAnesthesiology, 2006
- The Effect of the Interaction of Propofol and Alfentanil on Recall, Loss of Consciousness, and the Bispectral IndexAnesthesia & Analgesia, 1998
- The Effect of the Interaction of Propofol and Alfentanil on Recall, Loss of Consciousness, and the Bispectral IndexAnesthesia & Analgesia, 1998
- Influence of Age and Gender on the Pharmacokinetics and Pharmacodynamics of RemifentanilAnesthesiology, 1997
- Measuring the predictive performance of computer-controlled infusion pumpsJournal of Pharmacokinetics and Biopharmaceutics, 1992
- Closed-loop control of blood pressure, ventilation, and anesthesia deliveryJournal of Clinical Monitoring and Computing, 1987