Retrospective cohort study of the performance of the Pinnacle metal on metal (MoM) total hip replacement: a single-centre investigation in combination with the findings of a national retrieval centre
Open Access
Abstract
Objectives To determine risk factors for revision in patients implanted with a commonly used metal on metal (MoM) hip replacement. Design Retrospective cohort study in combination with a prospective national retrieval study (Northern Retrieval Registry (NRR)). Setting Combined orthopaedic unit in combination with the NRR. Participants All patients implanted with a DePuy Pinnacle MoM hip prostheses by the 2 senior authors were invited to attend for a review which included clinical examination, blood metal ion measurements, radiographs and targeted imaging. Explanted components underwent wear analysis using validated methodology and these results were compared with those obtained from the NRR. Results 489 MoM Pinnacle hips were implanted into 434 patients (243 females and 191 males). Of these, 352 patients attended the MoM recall clinics. 64 patients had died during the study period. For the purposes of survival analysis, non-attendees were assumed to have well-functioning prostheses. The mean follow-up of the cohort as a whole was 89 months. 71 hips were revised. Prosthetic survival for the whole cohort was 83.6% (79.9–87.3) at 9 years. The majority of explanted devices exhibited signs of taper junction failure. Risk factors for revision were bilateral MoM prostheses, smaller Pinnacle liners, and implantation in 2006 and later years. A significant number of devices were found to be manufactured out of their specifications. This was confirmed with analysis of the wider data set from the NRR. Conclusions This device was found to have an unacceptably high revision rate. Bilateral prostheses, those implanted into female patients and devices implanted in later years were found to be at greater risk. A significant number of explanted components were found to be manufactured with bearing diameters outside of the manufacturer's stated tolerances. Our findings highlight the clinical importance of hitherto unrecognised variations in device production.Keywords
This publication has 39 references indexed in Scilit:
- Five- to Ten-Year Outcomes for Modular Metal-On-Metal Total Hip ArthroplastyThe Journal of Arthroplasty, 2013
- Blood metal ion testing is an effective screening tool to identify poorly performing metal-on-metal bearing surfacesBone & Joint Research, 2013
- Deformation of the Durom Acetabular Component and Its Impact on Tribology in a Cadaveric Model—A Simulator StudyPLOS ONE, 2012
- Taper junction failure in large-diameter metal-on-metal bearingsBone & Joint Research, 2012
- Adverse reactions to metal debris: histopathological features of periprosthetic soft tissue reactions seen in association with failed metal on metal hip arthroplastiesJournal of Clinical Pathology, 2012
- Accelerating failure rate of the ASR total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Large Diameter Femoral Heads Impose Significant Alterations on the Strains Developed on Femoral Component and Bone: A Finite Element AnalysisThe Open Orthopaedics Journal, 2011
- Reduced Articular Surface of One-piece Cups: A Cause of Runaway Wear and Early FailureClinical Orthopaedics and Related Research, 2010
- Metal-on-Metal Hip Arthroplasty: Does Early Clinical Outcome Justify the Chance of an Adverse Local Tissue Reaction?Clinical Orthopaedics and Related Research, 2010
- 2008 John Charnley Award: Metal Ion Levels After Metal-on-Metal Total Hip Arthroplasty: A Randomized TrialClinical Orthopaedics and Related Research, 2009