Design and Characterization of a Sensorized Microfluidic Cell-Culture System with Electro-Thermal Micro-Pumps and Sensors for Cell Adhesion, Oxygen, and pH on a Glass Chip
Open Access
- 30 July 2015
- journal article
- research article
- Published by MDPI AG in Biosensors
- Vol. 5 (3) , 513-536
Abstract
We combined a multi-sensor glass-chip with a microfluidic channel grid for the characterization of cellular behavior. The grid was imprinted in poly-dimethyl-siloxane. Mouse-embryonal/fetal calvaria fibroblasts (MC3T3-E1) were used as a model system. Thin-film platinum (Pt) sensors for respiration (amperometric oxygen electrode), acidification (potentiometric pH electrodes) and cell adhesion (interdigitated-electrodes structures, IDES) allowed us to monitor cell-physiological parameters as well as the cell-spreading behavior. Two on-chip electro-thermal micro-pumps (ETμPs) permitted the induction of medium flow in the system, e.g., for medium mixing and drug delivery. The glass-wafer technology ensured the microscopic observability of the on-chip cell culture. Connecting Pt structures were passivated by a 1.2 μm layer of silicon nitride (Si3N4). Thin Si3N4 layers (20 nm or 60 nm) were used as the sensitive material of the pH electrodes. These electrodes showed a linear behavior in the pH range from 4 to 9, with a sensitivity of up to 39 mV per pH step. The oxygen sensors were circular Pt electrodes with a sensor area of 78.5 μm2. Their sensitivity was 100 pA per 1% oxygen increase in the range from 0% to 21% oxygen (air saturated). Two different IDES geometries with 30- and 50-μm finger spacings showed comparable sensitivities in detecting the proliferation rate of MC3T3 cells. These cells were cultured for 11 days in vitro to test the biocompatibility, microfluidics and electric sensors of our system under standard laboratory conditions.Keywords
This publication has 57 references indexed in Scilit:
- Experimental verification of an equivalent circuit for the characterization of electrothermal micropumps: High pumping velocities induced by the external inductance at driving voltages below 5 VElectrophoresis, 2012
- A short review on AC electro-thermal micropumps based on smeared structural polarizations in the presence of a temperature gradientColloids and Surfaces A: Physicochemical and Engineering Aspects, 2010
- A novel modular device for 3-D bone cell culture and non-destructive cell analysisActa Biomaterialia, 2010
- Modular glass chip system measuring the electric activity and adhesion of neuronal cells—application and drug testing with sodium valproic acidLab on a Chip, 2010
- ac-field-induced fluid pumping in microsystems with asymmetric temperature gradientsPhysical Review E, 2009
- Online monitoring of BALB/3T3 metabolism and adhesion with multiparametric chip-based systemAnalytical Biochemistry, 2007
- The third dimension bridges the gap between cell culture and live tissueNature Reviews Molecular Cell Biology, 2007
- Möglichkeiten und Grenzen der Mikrosensortechnologie in zellulärer Diagnostik und PharmascreeningChemie Ingenieur Technik, 2005
- A (Bio-)Chemical Field-Effect Sensor with Macroporous Si as Substrate Material and a SiO2 / LPCVD-Si3N4 Double Layer as pH TransducerSensors, 2002
- Oxygen diffusion in poly(dimethyl siloxane) using fluorescence quenching. I. Measurement technique and analysisJournal of Polymer Science Part A: Polymer Chemistry, 1986