Cyclic loading comparison of Bio‐SutureTak‐#2 FiberWire and Bio Mini‐Revo‐#2 Hi‐Fi suture anchor‐sutures in cadaveric scapulae
- 20 November 2007
- journal article
- research article
- Published by Wiley in Knee Surgery, Sports Traumatology, Arthroscopy
- Vol. 16 (3) , 317-325
Abstract
This study compared tap‐in Bio‐SutureTak suture anchor‐#2 FiberWire suture (Group 1) and screw‐in Bio Mini‐Revo suture anchor‐#2 Hi‐Fi suture (Group 2) fixation in the glenoid region of interest for Bankart repair, in addition to evaluation of isolated suture loop biomechanical properties under progressive incremental cyclic loads. With knowledge of glenoid apparent bone mineral density (BMD), implant preparation and fit characteristics, and following application of a light manual tensile load, the primary investigator scored each specimen for perceived within group biomechanical test performance using a 0–10 point modified visual analog scale. After scoring, 12 paired constructs were placed in a servo hydraulic device clamp, preloaded to 25 N, and cycled between 25 and 50 Hz with a 25 N load increase every 25 cycles. Group 2 withstood greater load (104.1 ± 56 vs. 70 ± 36.9 N, P = 0.04) and displaced more at failure (13 ± 4.5 vs. 8.6 ± 3.3 mm, P = 0.04). All Group 1 specimens failed prior to reaching 150 N, whereas 25% of Group 2 specimens (n = 3) failed at 200 N. All specimens failed by anchor pullout except for three Group 2 specimens that failed by eyelet breakage at 200 N. Isolated suture testing revealed that Group 1 sutures displaced less at each cyclic load (P = 0.028) and withstood greater failure loads (P = 0.028) than that of Group 2 sutures. Group 2 constructs displayed moderately strong relationships between perceived within group biomechanical test performance and ultimate load (r2 = 0.55) and displacement at failure (r2 = 0.67). Group 1 did not display significant relationships. Similar biomechanical performance between 50 and 125 N, greater load at failure, and superior biomechanical test prediction accuracy suggest that the screw‐in type Bio Mini‐Revo suture anchor‐#2 Hi‐Fi suture combination may be preferred for Bankart lesion repair in low apparent BMD glenoid processes. The #2 Hi‐Fi suture, however, allowed significantly greater displacement than the #2 FiberWire suture at each progressive cyclic load interval.Keywords
This publication has 39 references indexed in Scilit:
- Effects of a progressive loading exercise program on the bone and skeletal muscle properties of female osteopenic rats☆☆☆Experimental Gerontology, 2007
- Revision of Failed Arthroscopic Bankart RepairsThe American Journal of Sports Medicine, 2007
- The Bone Response to Non-Weight-Bearing Exercise Is Sport-, Site-, and Sex-SpecificClinical Journal of Sport Medicine, 2007
- Mechanical and Handling Properties of Braided Polyblend Polyethylene Sutures in Comparison to Braided Polyester and Monofilament Polydioxanone SuturesArthroscopy, 2006
- Sutures and Suture Anchors—Update 2006Arthroscopy, 2006
- Arthroscopic Bankart Repair With Knotless Suture Anchor for Traumatic Anterior Shoulder Instability: Results of Short‐Term Follow‐UpArthroscopy, 2006
- Practical light embalming technique for use in the surgical fresh tissue dissection laboratoryClinical Anatomy, 2005
- Biomechanical Evaluation of Different Suture Anchors for the Stabilization of Anterior Labrum LesionsArthroscopy, 2005
- Failure of anterior shoulder instability repair caused by eyelet cutout of absorbable suture anchorsArthroscopy, 2004
- Arthroscopic bankart reconstruction with a bioabsorbable anchorJournal of Shoulder and Elbow Surgery, 2003