RF power, B 1+rms , and SAR variation with RF coils, conductive metallic implants, and ionic solutions at 1.5T and 3T
Preprint
Abstract
Background and Purpose: The magnetic resonance imaging (MRI) access for patients with active and passive implants is limited by radiofrequency (RF) safety. The time-averaged root-mean-square RF field (B 1+rms ) and specific absorption rate (SAR) are being evaluated to monitor and control RF-induced heating near conductive metallic implants, such as deep brain stimulation (DBS) leads, during MRI. However, experimental methods to assess the relationship between RF power, B 1+rms, and SAR are lacking for RF coils, metallic implants, and ionic solutions. Materials and Methods: A method is developed to evaluate the variation of RF power, B 1+rms, and SAR with RF coils, metallic implants, and ionic solutions using phantoms consisting of water (H 2 O) and sodium chloride (NaCl) with four ionic concentrations (0, 1, 2, 3 %), four metallic wavelengths (0, λ /2, λ , 2 λ ), two RF coils (body, head) transmit/receive (Tx/Rx) combinations, and five RF pulse flip angles (30°, 45°, 60°, 75°, 90°) in two B 0 fields (1.5T and 3T). Results: The scanner-reported RF power and SAR varied with RF pulse sequences, RF coils, Tx/Rx, metallic implants, and ionic solutions, whereas B 1+rms varied only with RF pulse sequences. The RF power, B 1+rms , and SAR relationship depends on RF pulse sequences, RF coils, Tx/Rx, implant wavelengths, and ionic concentrations. SAR (whole-body, head) scaled with RF power by absorption ratios (α) variable with experimental conditions. Conclusions: B 1+rms is insensitive to the presence and absence of conductive metallic implants and ionic solutions, implant wavelengths, ionic concentrations, RF coils, and Tx/Rx combinations. RF power must be monitored because scanner-reported SAR may vary unpredictably with experiments.Keywords
This publication has 41 references indexed in Scilit:
- MRI‐Related Heating of Implants and Devices: A ReviewJournal of Magnetic Resonance Imaging, 2020
- Virtual population‐based assessment of the impact of 3 Tesla radiofrequency shimming and thermoregulation on safety and B1+ uniformityMagnetic Resonance in Medicine, 2015
- Reduction of implant RF heating through modification of transmit coil electric fieldMagnetic Resonance in Medicine, 2010
- Effect of the extracranial deep brain stimulation lead on radiofrequency heating at 9.4 Tesla (400.2 MHz)Journal of Magnetic Resonance Imaging, 2010
- EditorialNMR in Biomedicine, 2009
- Insight into RF power requirements and B1 field homogeneity for human MRI via rigorous FDTD approachJournal of Magnetic Resonance Imaging, 2007
- Deep brain stimulation for Parkinson's disease: Surgical issuesMovement Disorders, 2006
- Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 teslaJournal of Magnetic Resonance Imaging, 2002
- Neurostimulators: Potential for excessive heating of deep brain stimulation electrodes during magnetic resonance imagingJournal of Magnetic Resonance Imaging, 2001
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985