The objective of this study was to present 5 years of surgical experience, and the extended results of hearing preservation (based on 3-year follow-up), with the Med-El Vibrant Soundbridge (VSB) in which the floating mass transducer (FMT) is placed directly against the round window membrane, and the fascia is used only as covering tissue to keep it in position. A retrospective survey of surgical and audiological data was conducted to evaluate the performance and stability of patient hearing, with audiometric measurements performed over fixed time intervals up to 36 months. 21 patients, aged 19–62 years (mean 48.4), with mixed or conductive, bilateral or unilateral hearing loss were included in this study. Surgical intervention involved monaural implantation of the Med-El VSB between 2006 and 2009. The results were assessed using pure tone audiometry. In 5 years of experience with the technique, no significant complications or device extrusion were observed except for two revision surgeries requiring FMT repositioning. In the 3-year follow-up, we observed stable hearing in the implanted ear. It is concluded that direct round window stimulation without interposed fascia is an alternative for patients with hearing impairment caused by chronic otitis media and/or lack of ossicles, especially after modified radical mastoidectomy. It allows good results in a selected group of patients, although further observation on a larger population is needed to confirm long-term validity and effectiveness.Keywords
This publication has 5 references indexed in Scilit:
- Impact of Floating Mass Transducer Coupling and Positioning in Round Window VibroplastyOtology & Neurotology, 2011
- Analysis of Vibrant Soundbridge Placement Against the Round Window Membrane in a Human Cadaveric Temporal Bone ModelOtology & Neurotology, 2010
- Evaluation of Round Window Stimulation Using the Floating Mass Transducer by Intracochlear Sound Pressure Measurements in Human Temporal BonesOtology & Neurotology, 2010
- Factors Improving the Vibration Transfer of the Floating Mass Transducer at the Round WindowOtology & Neurotology, 2010
- Treatment of mixed hearing losses via implantation of a vibratory transducer on the round windowInternational Journal of Audiology, 2006