A New Redox Mediator (Cupric‐Neocuproine Complex)‐ Modified Pencil Graphite Electrode for the Electrocatalytic Oxidation of H 2 O 2 : A Flow Injection Amperometric Sensor
- 12 December 2019
- journal article
- research article
- Published by Wiley in ChemElectroChem
- Vol. 7 (3) , 649-658
Abstract
This paper introduces the cupric‐neocuproine complex ([Cu(Nc) 2 ] 2+ ) as a new redox mediator for the electrocatalytic oxidation of H 2 O 2 . For this purpose, [Cu(Nc) 2 ] 2+ was modified onto a disposable pencil graphite electrode (PGE) surface by using Nafion (Nf) as a cation‐exchange membrane for the effective adsorption of complex. The modified electrode was prepared by immersing Nf‐adsorbed PGE into [Cu(Nc) 2 ] 2+ complex solution (0.40 mM of Cu 2+ and 0.80 mM of Nc prepared at pH 4.76 in 0.10 M acetate buffer). SEM images and EDX spectra of electrodes were presented for the elucidation of their surface morphologies. Cyclic voltammetric results showed that H 2 O 2 was electrocatalytically oxidized at [Cu(Nc) 2 ] 2+ /Nf/PGE, because both the increase in current and the potential shift to a more negative direction were observed for oxidation of H 2 O 2 at [Cu(Nc) 2 ] 2+ /Nf/PGE in comparison to bare PGE. Then, the electrocatalytic activity of [Cu(Nc) 2 ] 2+ /Nf/PGE toward H 2 O 2 oxidation was utilized for amperometric determination of H 2 O 2 in flow injection analysis (FIA) system. Flow injection (FI) amperometric studies showed two linear calibration ranges (1.0‐1000.0 μM ( R 2 =0.9958) and 1000.0‐10000.0 μM ( R 2 =0.9902) with a detection limit of 0.4 μM H 2 O 2 which is considered good for electroanalytical determinations, especially for electrocatalytic oxidation. To test the applicability of the developed FI amperometric hydrogen peroxide sensor, H 2 O 2 contents of some relevant samples were determined.Keywords
This publication has 59 references indexed in Scilit:
- Novel Optical Fiber Reflectometric CUPRAC Sensor for Total Antioxidant Capacity Measurement of Food Extracts and Biological SamplesJournal of Agricultural and Food Chemistry, 2013
- Electrochemical study of PEDOT-PSS-MDB-modified electrode and its electrocatalytic sensing of hydrogen peroxideJournal of Solid State Electrochemistry, 2010
- Determination of antioxidants by a novel on-line HPLC-cupric reducing antioxidant capacity (CUPRAC) assay with post-column detectionAnalytica Chimica Acta, 2010
- Modified cupric reducing antioxidant capacity (CUPRAC) assay for measuring the antioxidant capacities of thiol-containing proteins in admixture with polyphenolsTalanta, 2009
- Measurement of xanthine oxidase inhibition activity of phenolics and flavonoids with a modified cupric reducing antioxidant capacity (CUPRAC) methodAnalytica Chimica Acta, 2009
- Hydroxyl radical scavenging assay of phenolics and flavonoids with a modified cupric reducing antioxidant capacity (CUPRAC) method using catalase for hydrogen peroxide degradationAnalytica Chimica Acta, 2008
- Spectrophotometric determination of ascorbic acid by the modified CUPRAC method with extractive separation of flavonoids–La(III) complexesAnalytica Chimica Acta, 2007
- DNA–Cu(II) poly(amine) complex membrane as novel catalytic layer for highly sensitive amperometric determination of hydrogen peroxideBiosensors and Bioelectronics, 2006
- Electrochemical reduction of oxygen and hydrogen peroxide catalyzed by a surface copper(II)–2,4,6-tris(2-piridil)-1,3,5-triazine complex adsorbed on a graphite electrodeJournal of Power Sources, 2005
- Electrocatalytic H2O2 amperometric detection using gold nanotube electrode ensemblesAnalytica Chimica Acta, 2004