Investigation of electrospray stability with a dynamic liquid flow splitter
- 24 December 2014
- journal article
- Published by Pleiades Publishing Ltd in Journal of Analytical Chemistry
- Vol. 69 (14) , 1320-1322
Abstract
The article gives the results of research of electrospray stability with the use of a dynamic splitting of a flow in the spray area under normal conditions. An acidified water-acetonitrile solution (1: 1) was used as an eluant. Experimental results for the used configuration of electrospray junction with a coaxial selection of fluid excess as a steam-gas mixture taken out of the dispersion area (meniscus fluid on the butt-end of a capillary) in the compass of a fluid flow along the capillary from 50 up to 200 μL/min showed more stable relation of an atomized solution flow in comparison with a mode without use of a dynamic splitting of an atomized fluid flow.Keywords
This publication has 2 references indexed in Scilit:
- Electrospray ion source with a dynamic liquid flow splitterRapid Communications in Mass Spectrometry, 2013