A cross-industry assessment of the flow rate-time profiles of test equipment typically used for dry-powder inhaler (DPI) testing: Part 1 – compendial apparatuses
- 14 August 2020
- journal article
- research article
- Published by Taylor & Francis in Aerosol Science and Technology
- Vol. 54 (12) , 1424-1447
Abstract
We report a cross-industry study characterizing flow rate-time profiles of equipment used for testing dry-powder inhalers (DPIs). Nine organizations used the same thermal mass flow sensor to record flow rate-time profiles at the inlet of individual participant compendial DPI test systems including either sample collection tubes (SCT), the Andersen 8-stage non-viable impactor (ACI) or the Next Generation Impactor (NGI™) with and without pre-separator (PS). The plan included some tests with a surrogate DPI consisting of one of three inlet orifices chosen to generate a 4-kPa pressure drop at each of the target final flow rates of 30, 60, and 90 l.min−1, simulating the pressure drop typical of high-, medium-, and low-resistance DPIs. When a particular surrogate DPI was present at the inlet, rise times to 90% of these target flow rates (t90) were shortest at the highest target flow rate, and decreased linearly with decreasing apparatus internal volume, following the order: NGI-PS > NGI > ACI-PS > ACI > SCT. A flow acceleration parameter was also evaluated, expressed as the slope between the rise times when the flow rate attained 20% and 80% of each final steady flow rate (slopet20/80). Flow acceleration was smallest at the lowest target flow rate, decreasing exponentially with increasing internal volume. Measurements were also made without the surrogate DPI, providing a reference condition with no resistance at the inlet to the apparatus. These flow rate-rise time profiles will be useful for those involved in evaluating equipment for characterizing DPIs and in understanding the behavior of these inhalers in development or commercial production.Keywords
This publication has 11 references indexed in Scilit:
- Internal Volumes of Pharmaceutical Compendial Induction Port, Next-Generation Impactor With and Without Its Pre-separator, and Several Configurations of the Andersen Cascade Impactor With and Without Pre-separatorJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2020
- Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler ® in Comparison with Diskus ® and Turbohaler ® Dry Powder InhalersJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2015
- Investigation of Dry Powder Inhaler (DPI) Resistance and Aerosol Dispersion Timing on Emitted Aerosol Aerodynamic Particle Sizing by Multistage Cascade Impactor when Sampled Volume Is Reduced from Compendial Value of 4 LAAPS PharmSciTech, 2014
- Effect of Device Design on the In Vitro Performance and Comparability for Capsule-Based Dry Powder InhalersThe AAPS Journal, 2012
- Effect of Sampling Volume on Dry Powder Inhaler (DPI)-Emitted Aerosol Aerodynamic Particle Size Distributions (APSDs) Measured by the Next-Generation Pharmaceutical Impactor (NGI) and the Andersen Eight-Stage Cascade Impactor (ACI)AAPS PharmSciTech, 2012
- Influence of Air Flow on the Performance of a Dry Powder Inhaler Using Computational and Experimental AnalysesPharmaceutical Research, 2005
- Cascade Impactors for the Size Characterization of Aerosols from Medical Inhalers: Their Uses and LimitationsJournal of Aerosol Medicine, 2003
- Effect of rise in simulated inspiratory flow rate and carrier particle size on powder emptying from dry powder inhalersAAPS PharmSci, 2000
- The Relationship Between Powder Inhaler Resistance and Peak Inspiratory Conditions in Healthy Volunteers — Implications for In Vitro TestingJournal of Aerosol Medicine, 1993
- Characteristics of laminar jet impactorsEnvironmental Science & Technology, 1974