  • The NGI Pre-Separator: Designed for Formulations with Large Carrier Particles

    by Daryl L. Roberts, Jolyon P. Mitchell

    Background: Compendial methods recommend including a pre-separator (PS) before the multi-stage impactor when testing dry powder inhalers (DPIs) for emitted aerosol aerodynamic particle size distribution (APSD). For data analysis, they are either silent or treat the PS as a “non-sizing” component. The overall message is “use the PS but ignore what it collects.” The Next Generation Impactor (NGI) PS reinforces this mindset because it was specifically designed for testing carrier-particle-based DPIs wherein any API in the PS results from a failure to detach from the large carrier particles. However, users with carrier-free DPI formulations might wonder what to do if drug is found in the PS and if it is technically justified to omit this component, thereby increasing laboratory throughput by relying on the NGI alone to provide accurate APSD metrics. Methods: Particle capture efficiency curves for the NGI PS were constructed from the published calibration data at flow rates of 15, 30, 60, and 100 L/min. These curves and those published for NGI stages 1–7 enabled calculation of how model input aerosols disperse throughout an NGI with or without a PS. APSD metrics were subsequently estimated from these calculated API mass fractions and then compared with and without the PS. Results: The presence of the PS yielded APSD metrics for the model aerosols that were almost indistinguishable from those obtained without the PS, accounting for the API mass recovered from the PS as if it had collected on an impactor stage. Conclusions: The PS is necessary when and only when the formulation contains large carrier particles. Consequently, for carrier-free formulations, method validation with no PS is likely to succeed. If for any reason it is retained in the test method of a carrier-free formulation, the PS should be treated like an impactor stage for APSD data analysis purposes.

  • Flow Control When Testing Dry-Powder Inhalers

    by Daryl L. Roberts

    Background: International compendia specify setting the flow for testing dry powder inhalers (DPIs) with the pressure ratio across the flow control valve ≤0.5. This guidance derives from thermodynamic theory, but there is no requirement that the flow control equipment demonstrates adherence to this theory. There is therefore a need to define and test for acceptable flow control valve performance, enabling DPI manufacturers to determine whether their test equipment is achieving the expected flow control. Herein, we present flow control data for two valves common in the inhaler testing community and describe the need for qualification specifications for flow control equipment with such valves. Methods: We devised a simple laboratory apparatus and method for determining the mass flow rate through two common flow control valves in their commercial configuration. In the test method, the pressure ratio across the valve is lowered from ∼0.7 to 0.3. This range of pressure ratio covers the standard test requirement that the pressure ratio must be 0.5 or less and maps the transition of compressible flow through the valve to the point where the air flow theoretically remains constant. Results: A common industrial valve, mentioned specifically in Pharm. Eur. 2.9.18 and in earlier editions of United States Pharmacopeia 601, exhibits nearly constant flow when the pressure ratio is <0.5. Another valve in widespread use exhibits a continued slight increase in mass flow rate even as the pressure ratio is decreased below 0.5. Conclusions: Inhaler testing laboratories should test whether their equipment exhibits the constant flow condition that is often assumed when the pressure ratio across the flow control valve is 0.5 or smaller. Quantitative specifications, yet to be developed, for qualifying commercial flow control equipment must account for both the transient start-up and the steady-state flow in DPI testing.

  • Multi-Participant Blinded Investigation into Internal Losses of Medication in Commercially Available Mixing Inlets: Testing Different Orally Inhaled Product Classes Following Pharmacopeial Methods

    by Patrik U. Andersson, Jim Clay, Mark Parry, Teresa Iley, Daryl L. Roberts, Lois Slator, Mårten Svensson, Hlack Mohammed, Jolyon P. Mitchell

    Background: The “Miller” design of mixing inlet (MI) enables a cascade impactor to operate at a constant flow rate while the orally inhaled product-on-test is evaluated at varying flow rates by controlling the flow of air via its side-arm. Study Purpose: As part of the European Pharmaceutical Aerosol Group (EPAG) Impactor subgroup, we report a cross-industry experimental investigation by five organizations to determine internal losses of different inhaler-generated aerosolized medications within commercially available MIs, focusing on pharmacopeial methods for product testing. Methods: Evaluations were undertaken of solution and suspension formulations delivered by pressurized metered dose inhalers (pMDIs), passive dry powder inhalers (DPIs), and compressed air-jet and vibrating mesh nebulizers. Four different apparatuses were evaluated at different constant air flow rates entering the MI side arm. The nebulizers were tested utilizing a variable adult flow profile generated by a breathing simulator. Results: Losses within the MI were generally <5%, expressed as a percentage of the delivered mass of active pharmaceutical ingredient (API) ex-inhaler. These losses were sufficiently small that they can in most cases be accommodated within the allowance of ±5% OIP label claim emitted mass/actuation in the pharmacopeial compendia for total internal losses for aerodynamic particle size distribution (APSD) determination. However, corresponding average losses were between 2.8% and 5.2% of the mass of API presented to the MI for the blister-based DPIs. APSD-derived measures were largely unaffected by the magnitude of pressurized air flow up to 60 L/min to the side-arm of the MI, except for the solution-formulated pMDI, where increasing flow rate was associated with reduced mass median aerodynamic diameter and increased geometric standard deviation, suggestive of a dependency related to ethanol co-solvent evaporation rate. Conclusions: MI loss evaluation should be considered an important part of method development to minimize internal losses of the aerosolized medication being sampled.

  • The Fundamentals of the Analysis of Cascade Impactor Data for the Characterization of Aerodynamic Particle Size Distributions (APSDs) in the Testing of Orally Inhaled Drug Products (OIPs)

    by Daryl L. Roberts, Jolyon P. Mitchell

    The multi-stage cascade impactor (CI) is the recognized apparatus for the characterization of the aerodynamic particle size distribution (APSD) of aerosols


  • Acceptance Testing of Used Cascade Impactor Stages Based on Pressure Drop Measurements in a Flow System Managed with a Critical Flow Venturi: Part I—Laboratory Proof-of-Principle

    by Daryl L. Roberts

    Background: Cascade impactors are essential for measuring the aerodynamic particle size distribution delivered by metered dose, dry powder, and similar inhalable drug products. For quality control of used impactors, periodic optical inspection of the nozzles of each impactor stage (stage mensuration) is currently the only method sufficiently precise to test whether used impactors are suitable for continued use, in accord with pharmacopeial standards. Here, we demonstrate a new method for quality control of used impactors. The method combines stage-wise pressure-drop measurement with a critical flow venturi (CFV) for air flow management. This technique avoids the unacceptably large uncertainty in conventional air flow rate measurements and instead relies on pressure and temperature measurement upstream of the CFV. These measurements can be made precisely with affordable equipment. Methods: We placed a toroidally shaped CFV downstream of a Next Generation Impactor™** (NGI) and precisely measured the stagnation pressure (±0.02%) and temperature (±0.03%) upstream of this CFV at impactor inlet flow rates close to 60 L/min. Pressure-drop measurements (±0.25%) at stages 3–7 and the micro-orifice collector were made with capacitive diaphragm transducers and with a special lid to the NGI that allowed pneumatic connection to the interstage passageways before and after each impactor stage. Results: The measured pressure drop values matched, to fractional percentage precision, those predicted by the incompressible flow theory through the nozzles and the compressible flow theory through the CFV. Conclusions: Practical equipment has been assembled that measures, to fractional percentage precision, the pressure drop through impactor nozzles at precisely managed flow conditions. The experimental results support the relevant flow principles. The results, thereby, support the use of this method for quantifying whether used impactor stages are suitable for continued use in the testing of registered inhalable drug products, in accord with pharmacopeial standards.


  • Acceptance Testing of Used Cascade Impactor Stages Based on Pressure Drop Measurements in a Flow System Managed with a Critical Flow Venturi: Part II—Quantification of the Test Uncertainty Ratio for Pass/Fail Decision-Making under Pharmacopeial Constraints

    by Daryl L. Roberts

    Background: The pressure drop at any cascade impactor stage is related to the open area of nozzles at that stage. Pressure drop measurement therefore can potentially test whether the nozzles of a given stage are within the range specified for continued use for testing of inhalable drug products. Previous such efforts, however, have been hindered by the measurement precision required for making a pass/fail decision about these used impactors. In this study, we articulate the error analysis for a pressure drop measurement system managed with a critical flow venturi (CFV) and show that the resultant uncertainty in the effective diameter of used Next Generation Impactor (NGI) and Andersen-type impactor stages is generally small compared to the specification range. This result enables the user to make a pass/fail decision regarding suitability for continued use. Methods: We develop the equations governing the relationship between stage pressure drop and the effective diameter of each stage of a used impactor. These equations show that pressure drop measurements can indicate only the change (if any) in the effective diameter between a previous measurement and the current measurement. Propagation-of-error principles therefore show that the uncertainty of both measurements affects the resulting uncertainty. Results: The test uncertainty ratio (analytical power) of a CFV-managed pressure drop measurement system exceeds six for all but stage one of the NGI and for stages −1 and −2 of the Andersen-type impactor. The stage-one nozzle of the NGI is readily qualified with a Class X pin. Conclusions: The CFV-managed flow system described in Part I is sufficiently precise to enable a decision to be made about whether used impactor nozzles are suitable for continued use for testing of registered inhalable drug products. Examination of the industrial viability of the technology will require long-term testing in real-world settings with comparison to optical inspection methods.


  • Pressure Drop Characteristics of the Reduced NGI Configuration with Several Common Glass Fiber Filters

    by Daryl L. Roberts

    Background: Measurement of aerodynamic particle size distribution, a clinically relevant in vitro attribute of inhalable drug products, involves multistage cascade impactors and is tedious and expensive. A leading candidate for a quicker method is the reduced NGI™ (rNGI). This method involves placing glass fiber filters on top of the nozzles of a chosen NGI stage, selected often to collect all particles with an aerodynamic diameter smaller than approximately five microns. These filters contribute additional flow resistance that can alter the flow rate start-up curve, potentially affecting the size distribution and mass of the drug product dispensed by passive dry powder inhalers (DPIs). The magnitude of these additional flow resistance measurements is currently unreported in the literature. Materials and Methods: We placed glass fiber filters on top of the stage 3 nozzles of an NGI, along with the necessary support screen and hold-down ring. We measured the pressure drop across NGI stage 3 with the assistance of a delta P lid and a high-precision pressure transducer. With each filter material type and multiple individual filters, we gathered eight replicates at flow rates of 30, 45, and 60 L/min. Results: The filters typically doubled the total pressure drop through the NGI. For example, at a flow rate of 60 L/min, the Whatman 934-AH filters introduced a pressure drop of about 9800 Pa at stage 3, reducing the absolute pressure exiting the NGI to about 23 kPa below ambient, compared with a typical value of 10 kPa for the NGI alone at this flow rate. Conclusions: The pressure drop across typical filters is approximately equal to that through the NGI alone and therefore will affect the flow start-up rate intrinsic to compendial testing of passive DPIs. This change in start-up rate could cause differences between results of the rNGI configuration and those of the full NGI and will increase the required vacuum pump capacity.


  • A cross-industry assessment of the flow rate-elapsed time profiles of test equipment typically used for dry-powder inhaler (DPI) testing: Part 2– analysis of transient air flow in the testing of DPIs with compendial cascade impactors

    by H. K. Versteeg, D. L. Roberts, F. Chambers, A. Cooper, M. Copley, J. P. Mitchell, H. Mohammed

    We describe a computational model that captures the physics of the unsteady air-flow start-up that accompanies the testing of dry-powder inhalers (DPI) with cascade impactors (CIs) specifically when following methods described in the pharmacopeial compendia. This transient has been measured in a multi-organization study for a wide range of conditions and reported in a companion article. The DPI test system is simulated as a series of flow resistances, starting with the inhaler itself, including each stage of the CI, and finishing with the prescribed flow control valve, wherein the flow is held at sonic velocity. The resulting non-dimensional equations indicate the relative importance of the several flow resistances. The model agrees well with the available experimental data for the Next Generation Impactor (NGI™) and qualitatively with the available data from a variety of the configurations of the Andersen 8-stage cascade impactor (ACI), including the typical 4-kPa pressure drop across an entry fixed orifice mimicking surrogate low-, medium- and high-resistance DPIs. The model indicates that the start-up times for the NGI and for the ACI are very reasonably estimated by a simple “reference time,” given in equation (26) and, for an inlet flow rate of 60 L/min, having a value of 277 ms and 113 ms for the NGI and ACI (60-L/min configuration), respectively. The model also enables a baseline, universal design curve for the flow rise-time performance of testing DPIs with the NGI (Figures 9a,b), because this impactor requires no change of components for any inlet flow rate.


  • A cross-industry assessment of the flow rate-time profiles of test equipment typically used for dry-powder inhaler (DPI) testing: Part 1 – compendial apparatuses

    by R. Greguletz, P. U. Andersson, A. Cooper, F. Chambers, M. A. Copley, G. Daniels, M. Hamilton, M. Hammond, H. Mohammed, D. L. Roberts, et al.

    We report a cross-industry study characterizing flow rate-time profiles of equipment used for testing dry-powder inhalers (DPIs). Nine organizations used the same thermal mass flow sensor to record flow rate-time profiles at the inlet of individual participant compendial DPI test systems including either sample collection tubes (SCT), the Andersen 8-stage non-viable impactor (ACI) or the Next Generation Impactor (NGI™) with and without pre-separator (PS). The plan included some tests with a surrogate DPI consisting of one of three inlet orifices chosen to generate a 4-kPa pressure drop at each of the target final flow rates of 30, 60, and 90 l.min−1, simulating the pressure drop typical of high-, medium-, and low-resistance DPIs. When a particular surrogate DPI was present at the inlet, rise times to 90% of these target flow rates (t90) were shortest at the highest target flow rate, and decreased linearly with decreasing apparatus internal volume, following the order: NGI-PS > NGI > ACI-PS > ACI > SCT. A flow acceleration parameter was also evaluated, expressed as the slope between the rise times when the flow rate attained 20% and 80% of each final steady flow rate (slopet20/80). Flow acceleration was smallest at the lowest target flow rate, decreasing exponentially with increasing internal volume. Measurements were also made without the surrogate DPI, providing a reference condition with no resistance at the inlet to the apparatus. These flow rate-rise time profiles will be useful for those involved in evaluating equipment for characterizing DPIs and in understanding the behavior of these inhalers in development or commercial production.