Increased Radiation Dose to Overweight and Obese Patients from Radiographic Examinations
Authors:
- Jacquelyn C. YanchJYJacquelyn C. Yanch
- Richard H. BehrmanRBRichard H. Behrman
- Michael J. HendricksMHMichael J. Hendricks
- John H. McCallJMJohn H. McCall
Open Access
- 1 July 2009
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 252 (1) , 128-139
Abstract
To estimate the increase in effective radiation dose from diagnostic x-rays for overweight and obese adult patients, as compared with the effective dose for lean reference phantoms. Relative effective radiation doses (E/E(0)) for the acquisition of chest and abdominal radiographs were calculated by using Monte Carlo computer simulations of effective doses delivered to adult phantoms with (E) and without (E(0)) subcutaneous adipose tissue added to the torso for five fat distributions. Total (anterior plus posterior) fat thicknesses ranged from 0 to 38 cm. For 30 cm of additional fat, E/E(0) values for 120-kVp chest and 80-kVp abdomen radiographs ranged from approximately 2 to 31 and 2 to 83 for male patients, respectively, and from 2 to 45 and 2 to 76 for female patients, respectively, depending on the type of fat distribution and patient orientation in the x-ray beam (anteroposterior or posteroanterior). Orienting the patient such that the thinnest fat layer was facing away from the x-ray tube minimized E/E(0), which was well approximated by using the formula E/E(0) = [B(t)/B(0)] x exp(kt(DF)), where B(t) and B(0) are the antiscatter grid Bucky factors for patient thicknesses of t and t = 20 cm, respectively; k, a constant; and t(DF), the distal (beam exit) fat layer thickness. Reductions in E/E(0) reached 14% and 20% for the thickest phantoms when x-ray tube voltages were increased by 10 and 20 kVp, respectively, for abdominal radiography in the male phantom. Effective doses from radiographic examinations in the extremely obese can exceed 100 mSv from only a small number of abdominal examinations and should be minimized to the extent possible and monitored. Exponential dose increases for increased subcutaneous fat thicknesses can be reduced substantially by positioning the patient so that the thinnest fat layer (anterior or posterior) is closest to the image receptor. Increasing the tube voltage also reduces the dose-but to a much smaller extent.Keywords
This publication has 10 references indexed in Scilit:
- MEDICAL RADIATION EXPOSURE IN THE U.S. IN 2006: PRELIMINARY RESULTSHealth Physics, 2008
- Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systemsPhysics in Medicine & Biology, 2007
- Obesity's Role in Heart Disease Requires Apples and Pears ComparisonJAMA, 2005
- A review of US anthropometric reference data (1971–2000) with comparisons to both stylized and tomographic anatomic modelsPhysics in Medicine & Biology, 2003
- Effect of radiographic techniques (kVp and mAs) on image quality and patient doses in digital subtraction angiographyMedical Physics, 2002
- Obesity, Health Services Use, and Health Care Costs Among Members of a Health Maintenance OrganizationArchives of internal medicine (1960), 1998
- Overweight and obesity in the United States: prevalence and trends, 1960–1994International Journal of Obesity, 1997
- Computing patient doses of X-ray examinations using a patient size- and sex-adjustable phantom.British Journal of Radiology, 1997
- Revised radiation doses for typical X-ray examinations. Report on a recent review of doses to patients from medical X-ray examinations in the UK by NRPB. National Radiological Protection Board.British Journal of Radiology, 1997
- IntroductionAnnals of the ICRP, 1991