Abstract
A sensitive capillary electrophoretic method featuring spectrophotometric detection using a commercial Z‐cell was devised for the assay of 8‐hydroxy‐2′‐deoxyguanosine (8OHdG) in human urine. Solid‐phase extraction (SPE) based on hydrophilic‐lipophilic‐balanced RP sorbent was utilized for urine sample pretreatment and analyte preconcentration. The separation was carried out in conventional fused‐silica capillaries employing a Z‐cell with hydrodynamic sample injection (at 50 mbar for 12 s). The BGE (pH* 9.2, adjusted with 1 M NaOH) contained 0.15 M boric acid and 10% v/v ACN. The detection wavelength was 282 nm. The calibration curve for 8OHdG (measured in spiked urine) was linear in the range 10–1000 ng/mL; R2 = 0.9993. The LOD was 3 ng/mL (11 nmol/L) of 8OHdG. Determination of the 8OHdG urinary levels was possible even in healthy individuals.
Funding Information
  • Charles University Grant Agency (GAUK 263 811)
  • European Social Fund and the state budget of the Czech Republic (CZ.1.07/2.3.00/30.0061)

This publication has 18 references indexed in Scilit: