Doses to patients from routine diagnostic X-ray examinations in England
Authors:
- P. C. ShrimptonPSP. C. Shrimpton
- B. F. WallBWB. F. Wall
- D. G. JonesDJD. G. Jones
- E. S. FisherEFE. S. Fisher
- M. C. HillierMHM. C. Hillier
- G. M. KendallGKG. M. Kendall
- R. M. HarrisonRHR. M. Harrison
- 1 August 1986
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 59 (704) , 749-758
Abstract
A collaborative survey between the National Radiological Protection Board and the Hospital Physicists' Association has been conducted to ascertain current levels of exposure for patients undergoing 10 routine types of X-ray examination in England. The main part of this study consisted of measurements on nearly 3200 patients attending 20 randomly selected English hospitals. The energy imparted to each patient was determined from a measurement of the total exposure–area product for the examination. In addition, thermoluminescent dosemeters were attached to the patient's skin to enable the derivation of doses to the major radiosensitive organs, either directly or using appropriate conversion factors calculated for a mathematical phantom by a Monte Carlo technique. Histograms are presented showing the wide distributions often observed in the doses for each type of examination. Mean values of exposure–area product, energy imparted to the patient, entrance skin dose per film and organ dose are reported, togethe...Keywords
This publication has 7 references indexed in Scilit:
- Patient Exposure Trends in Medical and Dental RadiographyHealth Physics, 1986
- Calculation of patient skin dose from diagnostic X-ray proceduresBritish Journal of Radiology, 1985
- Reduction of fluoroscopic exposure for the air-contrast barium enemaBritish Journal of Radiology, 1983
- A survey of radiation doses to patients in five common diagnostic examinations. I. Principles and techniques of dose evaluation by indirect measurement II. Survey resultsBritish Journal of Radiology, 1983
- An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinationsPhysics in Medicine & Biology, 1982
- Energy Imparted in Roentgen Diagnostic ProceduresActa Radiologica. Diagnosis, 1979
- Patient Exposures and Radiation Risks In Swedish Diagnostic RadiologyActa Radiologica: Oncology, Radiation, Physics, Biology, 1978