Comparison of Ocular Wavefront in Seated and Supine Positions Using a Hand-Held Hartmann–Shack Aberrometer
Open Access
- 22 September 2025
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 14 (18) , 6688
Abstract
Background/Objectives: Intraoperative aberrometry has gained clinical relevance in correcting aberrations during cataract and corneal refractive surgeries. As wavefront aberrations are typically measured with patients seated, while surgeries are performed supine, this study aimed to compare ocular aberrations between seated and supine positions, using a hand-held Hartmann–Shack aberrometer. Methods: Total ocular wavefront aberrations were measured five times consecutively from cyclopledged eyes for a 6 mm pupil, under three conditions: (1) seated with a holder-fixed mode (reference), (2) seated with a hand-held mode, and (3) supine with a hand-held mode. Condition 2 was included to assess potential measurement errors from the hand-held mode. Repeatability was assessed using the standard deviation of repeated measurements (Srm) and the intraclass correlation coefficient (ICC). Differences among the three conditions were analyzed to evaluate the effects of positional change and hand-held stability on ocular wavefront aberration measurements. Results: Eighteen healthy subjects (36 eyes) were enrolled. The Srms for the sphere were 0.37, 0.38, and 0.40 diopters (D); and for the cylinder, 0.10, 0.10, and 0.11 D, with no significant differences across conditions. ICC values exceeded 0.9 for both lower-order aberrations (LOAs) and higher-order aberrations (HOAs), indicating excellent repeatability. The mean root mean square HOAs (HOA_RMS) were 0.48, 0.49, and 0.45 µm, with no statistically significant differences by position (p = 0.913) or measurement mode (p = 0.966). Conclusions: The hand-held Hartmann–Shack aberrometer demonstrated satisfactory repeatability for LOAs and HOAs. Supine measurements did not differ from seated, supporting the feasibility of extending preoperative results to intraoperative aberrometry, regardless of positional differences.Keywords
Funding Information
- National Research Foundation of Korea (NRF) (IRIS RS-2024-00457444)
This publication has 61 references indexed in Scilit:
- Intraocular Lens Calculations2024
- Controlled elevation of intraocular pressure and its impact on ocular aberrations in healthy eyesExperimental Eye Research, 2018
- Effect of Different Head Positions in Lateral Decubitus Posture on Intraocular Pressure in Treated Patients With Open-Angle GlaucomaAmerican Journal of Ophthalmology, 2015
- The Effect of Lateral Decubitus Position on Intraocular Pressure in Patients with Untreated Open-Angle GlaucomaAmerican Journal of Ophthalmology, 2013
- Precision of higher-order aberration measurements with a new Placido-disk topographer and Hartmann-Shack wavefront sensorJournal of Cataract & Refractive Surgery, 2013
- Changes in the tear film and ocular surface after cataract surgeryJapanese Journal of Ophthalmology, 2012
- Detection of static cyclotorsion and compensation for dynamic cyclotorsion in laser in situ keratomileusisJournal of Cataract & Refractive Surgery, 2010
- Changes of Corneal Aberrations in Sitting and Supine PositionsAmerican Journal of Ophthalmology, 2006
- Cyclotorsional eye motion occurring between wavefront measurement and refractive surgeryJournal of Cataract & Refractive Surgery, 2004
- Torsional vestibulo-ocular reflex during whole-body oscillation in the upright and the supine position.Experimental Brain Research, 2000