A patient-specific planning target volume used in 'plan of the day' adaptation for interfractional motion mitigation
Authors:
Open Access
- 1 July 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Radiation Research
- Vol. 54 (suppl 1) , i82-i90
Abstract
We propose a patient-specific planning target volume (PTV) to deal with interfractional variations, and test its feasibility in a retrospective treatment-planning study. Instead of using one planning image only, multiple scans are taken on different days. The target and organs at risk (OARs) are delineated on each images. The proposed PTV is generated from a union of those target contours on the planning images, excluding voxels of the OARs, and is denoted the PTV ‘GP–OAR’ (global prostate–organs at risk). The study is performed using ‘plan of the day’ adaptive workflow, which selects a daily plan from a library of plans based on a similarity comparison between the daily scan and planning images. The daily plans optimized for GP–OAR volumes are compared with those optimized for PTVs generated from a single prostate contour (PTV SP). Four CT serials of prostate cancer patient datasets are included in the test, and in total 28 fractions are simulated. The results show that the daily chosen GP–OAR plans provide excellent target coverage, with V95 values of the prostate mostly > 95%. In addition, dose delivered to the OARs as calculated from applying daily chosen GP–OAR plans is slightly increased but comparable to that calculated from applying daily SP plans. In general, the PTV GP–OARs are able to cover possible target variations while keeping dose delivered to the OARs at a similar level to that of the PTV SPs.Keywords
This publication has 21 references indexed in Scilit:
- A Review of Update Clinical Results of Carbon Ion RadiotherapyJapanese Journal of Clinical Oncology, 2012
- Particle therapy at the Heidelberg Ion Therapy Center (HIT) – Integrated research-driven university-hospital-based radiation oncology service in Heidelberg, GermanyRadiotherapy and Oncology, 2010
- Heavy-ion tumor therapy: Physical and radiobiological benefitsReviews of Modern Physics, 2010
- Respiratory motion management in particle therapyMedical Physics, 2010
- Charged particles in radiation oncologyNature Reviews Clinical Oncology, 2009
- Target motion tracking with a scanned particle beamMedical Physics, 2007
- Accounting for range uncertainties in the optimization of intensity modulated proton therapyPhysics in Medicine & Biology, 2007
- Particle Radiation Therapy Using Proton and Heavier Ion BeamsJournal of Clinical Oncology, 2007
- Radiation Therapy With Charged ParticlesSeminars in Radiation Oncology, 2006
- Overview of clinical experiences on carbon ion radiotherapy at NIRSRadiotherapy and Oncology, 2004