Couch-based motion compensation: modelling, simulation and real-time experiments
Authors:
- Olivier C L HaasOlivier C L Haas
- Piotr SkworcowPSPiotr Skworcow
- Daniel PaluszczyszynDPDaniel Paluszczyszyn
- Abdelhamid SahihASAbdelhamid Sahih
- Mariusz RutaMRMariusz Ruta
- John A MillsJMJohn A Mills
- 5 September 2012
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (18) , 5787-5807
Abstract
The paper presents a couch-based active motion compensation strategy evaluated in simulation and validated experimentally using both a research and a clinical Elekta Precise Table™. The control strategy combines a Kalman filter to predict the surrogate motion used as a reference by a linear model predictive controller with the control action calculation based on estimated position and velocity feedback provided by an observer as well as predicted couch position and velocity using a linearized state space model. An inversion technique is used to compensate for the dead-zone nonlinearity. New generic couch models are presented and applied to model the Elekta Precise Table™ dynamics and nonlinearities including dead zone. Couch deflection was measured for different manufacturers and found to be up to 25 mm. A feed-forward approach is proposed to compensate for such couch deflection. Simultaneous motion compensation for longitudinal, lateral and vertical motions was evaluated using arbitrary trajectories generated from sensors or loaded from files. Tracking errors were between 0.5 and 2 mm RMS. A dosimetric evaluation of the motion compensation was done using a sinusoidal waveform. No notable differences were observed between films obtained for a fixed- or motion-compensated target. Further dosimetric improvement could be made by combining gating, based on tracking error together with beam on/off time, and PSS compensation.Keywords
This publication has 35 references indexed in Scilit:
- Towards MRI-guided linear accelerator control: gating on an MRI acceleratorPhysics in Medicine & Biology, 2011
- A robotic approach to 4D real-time tumor tracking for radiotherapyPhysics in Medicine & Biology, 2011
- Standardized evaluation of simultaneous integrated boost plans on volumetric modulated arc therapyPhysics in Medicine & Biology, 2010
- Incorporating system latency associated with real-time target tracking radiotherapy in the dose prediction stepPhysics in Medicine & Biology, 2010
- Real‐time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLCMedical Physics, 2010
- Strategies for Online Organ Motion Correction for Intensity-Modulated Radiotherapy of Prostate Cancer: Prostate, Rectum, and Bladder Dose EffectsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Investigation of motion sickness and inertial stability on a moving couch for intra-fraction motion compensationActa Oncologica, 2009
- Tumor tracking and motion compensation with an adaptive tumor tracking system (ATTS): System description and prototype testingMedical Physics, 2008
- A multiple model approach to respiratory motion prediction for real-time IGRTPhysics in Medicine & Biology, 2008
- Inferential modeling and predictive feedback control in real-time motion compensation using the treatment couch during radiotherapyPhysics in Medicine & Biology, 2007