Abstract
In the context of point of care testing (PoCT) and ISO 22870, internal quality control (IQC) is a crucial part of PoCT accreditation processes. Quality Control materials shall be periodically examined with a frequency that is based on the robustness of the analytical procedure and the risk of harm to the patient from an erroneous result. We propose to apply the statistical quality control (SQC) procedure to develop an individualized QC plan for AQT90 flex instrument used in PoCT. The robustness is determined by the sigma-metric and analytical goal represented by an allowable total error (TEa) is evaluated using a Varela graphic tool. A Sigma-metric SQC run size nomogram for estimating the number of patient samples between IQC events. According to the calculated robustness we can distinguish 3 groups of parameters: HCG and CRP with large sample size per event, D-Dimer and Procalcitonin with an average sample size per event and Myoglobin. NT-proBNP. and Troponin T with a limited sample size per event. In PoCT, the SQC strategy can promote more effective, and not necessarily more frequent, IQC.Keywords
This publication has 11 references indexed in Scilit:
- Risk analysis and assessment based on Sigma metrics and intended useBiochemia Medica, 2018
- Comprehensive evaluation of the internal and external quality control to redefine analytical quality goalsBiochemia Medica, 2018
- Selecting Statistical Procedures for Quality Control Planning Based on Risk ManagementClinical Chemistry, 2016
- Accreditation process in European countries – an EFLM surveyClinical Chemistry and Laboratory Medicine (CCLM), 2016
- Internal quality control in point-of-care testing: where’s the evidence?Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2015
- Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory MedicineClinical Chemistry and Laboratory Medicine (CCLM), 2015
- Assessing quality on the Sigma scale from proficiency testing and external quality assessment surveysClinical Chemistry and Laboratory Medicine (CCLM), 2015
- Assessing the Impact of the Frequency of Quality Control Testing on the Quality of Reported Patient ResultsClinical Chemistry, 2008
- Biological Variation of Procalcitonin in Healthy IndividualsClinical Chemistry, 2004
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999