Compact bunker shielding assessment for 1.5 T MR-Linac
Authors:
Open Access
- 25 April 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-9
Abstract
This study evaluated the effect of the 1.5 T magnetic field of the magnetic resonance-guided linear accelerator (MR-Linac) on the radiation leakage doses penetrating the bunker radiation shielding wall. The evaluated 1.5 T MR-Linac Unity system has a bunker of the minimum recommended size. Unlike a conventional Linac, both primary beam transmission and secondary beam leakage were considered independently in the design and defined at the machine boundary away from the isocenter. Moreover, additional shielding was designed considering the numerous ducts between the treatment room and other rooms. The Linac shielding was evaluated by measuring the leakage doses at several locations. The intrinsic vibration and magnetic field were inspected at the proposed isocenter of the system. For verification, leakage doses were measured before and after applying the magnetic field. The intrinsic vibration and magnetic field readings were below the permitted limit. The leakage dose (0.05–12.2 µSv/week) also complied with internationally stipulated limits. The special shielding achieved a five-fold reduction in leakage dose. Applying the magnetic field increased the leakage dose by 0.12 to 4.56 µSv/week in several measurement points, although these values fall within experimental uncertainty. Thus, the effect of the magnetic field on the leakage dose could not be ascertained.Keywords
Funding Information
- National Research Foundation of Korea (2019R1I1A1A01062157)
- Research Grant of Gangnam Severance Hospital (DHHD000067)
- National Research Foundation of Korea (2022R1A2C2011556)
This publication has 24 references indexed in Scilit:
- A formalism for reference dosimetry in photon beams in the presence of a magnetic fieldPhysics in Medicine & Biology, 2018
- Basic principles of magnetic resonance imaging for beginner oral and maxillofacial radiologistsOral Radiology, 2017
- Reference dosimetry in magnetic fields: formalism and ionization chamber correction factorsMedical Physics, 2016
- Assessment of computational tools for MRI RF dosimetry by comparison with measurements on a laboratory phantomPhysics in Medicine & Biology, 2015
- General Principles of MRI2013
- Reducing radiation exposure to patients from kV-CBCT imagingRadiotherapy and Oncology, 2010
- Radio frequency shielding for a linac-MRI systemPhysics in Medicine & Biology, 2010
- Study design in fMRI: Basic principlesBrain and Cognition, 2006
- A Review of Applications of MRI in Soft Tissue and Bone TumorsOrthopedics, 1988
- MRI and CT evaluation of primary bone and soft-tissue tumorsAmerican Journal of Roentgenology, 1986