A formalism for reference dosimetry in photon beams in the presence of a magnetic field
Open Access
- 1 June 2018
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 63 (12) , 125008
Abstract
A generic formalism is proposed for reference dosimetry in the presence of a magnetic field. Besides the regular correction factors from the conventional reference dosimetry formalisms, two factors are used to take into account magnetic field effects: (1) a dose conversion factor to correct for the change in local dose distribution and (2) a correction of the reading of the dosimeter used for the reference dosimetry measurements. The formalism was applied to the Elekta MRI-Linac, for which the 1.5 T magnetic field is orthogonal to the 7 MV photon beam. For this setup at reference conditions it was shown that the dose decreases with increasing magnetic field strength. The reduction in local dose for a 1.5 T transverse field, compared to no field is 0.51% ± 0.03% at the reference point of 10 cm depth. The effect of the magnetic field on the reading of the dosimeter was measured for two waterproof ionization chambers types (PTW 30013 and IBA FC65-G) before and after multiple ramp-up and ramp-downs of the magnetic field. The chambers were aligned perpendicular and parallel to the magnetic field. The corrections of the readings of the perpendicularly aligned chambers were 0.967 ± 0.002 and 0.957 ± 0.002 for respectively the PTW and IBA ionization chambers. In the parallel alignment the corrections were small; 0.997 ± 0.001 and 1.002 ± 0.003 for the PTW and IBA chamber respectively. The change in reading due to the magnetic field can be measured by individual departments. The proposed formalism can be used to determine the correction factors needed to establish the absorbed dose in a magnetic field. It requires Monte Carlo simulations of the local dose and measurements of the response of the dosimeter. The formalism was successfully implemented for the MRI-Linac and is applicable for other field strengths and geometries.Keywords
Funding Information
- EURAMET (EMPIR grant 15HLT08 MRgRT)
This publication has 28 references indexed in Scilit:
- A water calorimeter for on-site absorbed dose to water calibrations in60Co and MV-photon beams including MRI incorporated treatment equipmentPhysics in Medicine & Biology, 2016
- Consequences of air around an ionization chamber: Are existing solid phantoms suitable for reference dosimetry on an MR‐linac?Medical Physics, 2016
- TH-CD-BRA-05: First Water Calorimetric Dw Measurement and Direct Measurement of Magnetic Field Correction Factors, KQ,B, in a 1.5 T B-Field of An MRI LinacMedical Physics, 2016
- Evaluation of a commercial MRI Linac based Monte Carlo dose calculation algorithm with geant 4Medical Physics, 2016
- The Australian Magnetic Resonance Imaging–Linac ProgramSeminars in Radiation Oncology, 2014
- The Rotating Biplanar Linac–Magnetic Resonance Imaging SystemSeminars in Radiation Oncology, 2014
- Fast dose calculation in magnetic fields withGPUMCDPhysics in Medicine & Biology, 2011
- GPUMCD: A new GPU‐oriented Monte Carlo dose calculation platformMedical Physics, 2011
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999