A model based DC analysis of SiPM breakdown voltages

Authors:
  • FNFerenc Nagy
    Ferenc Nagy
  • GHGyula Hegyesi
    Gyula Hegyesi
  • GKGabor Kalinka
    Gabor Kalinka
  • JMJozsef Molnar
    Jozsef Molnar

Abstract
A new method to determine the breakdown voltage of SiPMs is presented. It is backed up by a DC model which describes the breakdown phenomenon by distinct avalanche turn-on ($V_{01}$)and turn off ($V_{10}$) voltages. It is shown that $V_{01}$ is related to the 'breakdown voltage' that previous DC methods derive from simple reverse current-voltage measurements, while $V_{10}$ is the 'real' breakdown voltage commonly obtained from complex gain-voltage measurements. The proposed method reveals how the microcell population distributes around $V_{01}$. It is found that if this distribution is assumed to be normal, then both $V_{01}$ and $V_{10}$ of the SiPM can readily be extracted from current-voltage curves. Measurements are in good agreement with the theoretical model.

This publication has 8 references indexed in Scilit: