Characterization Measurements Methodology and Instrumental Set-Up Optimization for New SiPM Detectors - Part II: Optical Tests
- 4 June 2014
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 14 (10) , 3567-3578
Abstract
A comprehensive and in-depth characterization procedure for obtaining very accurate measurements on silicon photomultiplier detectors (SiPMs) is here described. A large amount of optical tests are systematically carried out and discussed in terms of the most important SiPM performance parameters; in particular, an accurate estimation of the photon detection efficiency in the 350-900-nm wavelength spectral range and in steps of 10 nm is achieved, based on the single-photon counting technique, with substraction of the dark noise contribution and avoiding the additional noise sources of crosstalk and afterpulsing. Some recently produced detectors are analyzed and their relevant electro-optical parameters are evaluated in order to demonstrate the effectiveness and efficacy of the adopted characterization procedure and data-handling protocols in assessing the overall SiPM performance, regardless of the specific device tested. Tests repeatibility is carefully verified and all the evaluated parameter trends are proved to be compatible with the physics theory of the SiPM device.Keywords
This publication has 20 references indexed in Scilit:
- Accurate Analytical Single-Photoelectron Response of Silicon PhotomultipliersIEEE Sensors Journal, 2014
- Over saturation behavior of SiPMs at high photon exposureNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- Improved SPICE electrical model of silicon photomultipliersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Characterization and simulation of the response of Multi-Pixel Photon Counters to low light levelsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Precision measurements of Photon Detection Efficiency for SiPM detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009
- Characterization of detectors for the Italian Astronomical Quantum Photometer ProjectJournal of Modern Optics, 2009
- Characterization of SPAD Arrays: First Results2006
- Preliminary Test Measurements of the SPAD Array2004
- Silicon planar technology for single-photon optical detectorsIEEE Transactions on Electron Devices, 2003
- Characterization of a photon-counting intensified active pixel sensor (PC-IAPS): preliminary results2001