Automated phosphor thermometry lifetime calibration of multiple phosphors and emission lines to above 1900 K

Authors:

Abstract
A method is specified which enables lifetime calibration of multiple phosphors and emission lines at the same time to temperatures above 1900 K. The experimental setup and algorithm used for data collection and experimental equipment control are described. The phosphors were coated on an alumina oxide disc and the reference temperature was measured using three type B thermocouples. The algorithm automates the data collection process such that no input from an operator is required during operation. The potential systematic error in calibration temperature was evaluated and was less than 1% around 1400 K.
Funding Information
  • Energimyndigheten (45400-01)
  • European Commission (HYFLEXPOWER 884229)

This publication has 16 references indexed in Scilit: