3D MR elastography at 0. 55 T : Concomitant field effects and feasibility
Open Access
- 25 November 2024
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 93 (4) , 1602-1614
Abstract
Purpose: To demonstrate the feasibility of hepatic 3D MR elastography (MRE) at 0.55 T in healthy volunteers using Hadamard encoding and to study the effects of concomitant fields in the domain of MRE in general. Methods: Concomitant field effects in MRE are assessed using a Taylor series expansion and an encoding scheme is proposed to study the corresponding effects on 3D MRE at 0.55 T in numerical simulations and in phantom experiments. In addition, five healthy volunteers were enrolled and scanned at 60 Hz mechanical excitation with a Hadamard‐encoded 3D MRE sequence at 0.55 T and were also scanned with a reference 3D MRE sequence at 3 T for comparison. The retrieved biomechanical parameters were the magnitude of the complex shear modulus (| G *|), the shear wave speed (Cs), and the loss modulus ( G ″). Comparison of apparent SNR between 3 T and 0.55 T was performed. Results: Theoretical analysis, numerical simulations and phantom experiments demonstrated that the effects of concomitant fields in 3D MRE at 0.55 T are negligible. In the healthy volunteer experiments, the mean values of | G *|, Cs, and G″ in the liver were 2.1 ± 0.3 kPa, 1.5 ± 0.1 m/s, and 0.8 ± 0.1 kPa at 0.55 T, respectively, and 2.0 ± 0.2 kPa, 1.5 ± 0.1 m/s, and 0.9 ± 0.1 kPa at 3 T, respectively. Bland–Altman analysis demonstrated good agreement between the biomechanical parameters retrieved at 0.55 T and 3 T. A 2.1‐fold relative apparent SNR decrease was observed in 3D MRE at 0.55 T in comparison with 3 T. Conclusion: Hepatic 3D MRE is feasible at 0.55 T, showing promising initial results in healthy volunteers.Keywords
Funding Information
- Cancer Research UK (C7893/A26233)
- Siemens Healthineers
- Wellcome EPSRC Centre for Medical Engineering (WT203148/Z/16/Z)
This publication has 41 references indexed in Scilit:
- Single Breath‐Hold MR Elastography for Fast Biomechanical Probing of Pancreatic StiffnessJournal of Magnetic Resonance Imaging, 2023
- Correlated noise in brain magnetic resonance elastographyMagnetic Resonance in Medicine, 2021
- A compact 0.5 T MR elastography device and its application for studying viscoelasticity changes in biological tissues during progressive formalin fixationMagnetic Resonance in Medicine, 2017
- Interplatform reproducibility of liver and spleen stiffness measured with MR elastographyJournal of Magnetic Resonance Imaging, 2015
- Magnetic Resonance Imaging2014
- MR elastography of healthy liver parenchyma: Normal value and reliability of the liver stiffness value measurementJournal of Magnetic Resonance Imaging, 2012
- Review of MR elastography applications and recent developmentsJournal of Magnetic Resonance Imaging, 2012
- Obesity and Nonalcoholic Fatty Liver Disease: Biochemical, Metabolic, and Clinical ImplicationsHepatology, 2010
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Concomitant gradient terms in phase contrast MR: Analysis and correctionMagnetic Resonance in Medicine, 1998